Fabrication of Small Caliber Stent-grafts Using Electrospinning and Balloon Expandable Bare Metal Stents
Published on: October 26, 2016
Feasibility and safety of duodenal covered self-expandable metallic stent fixation: an experimental study
Yasuki Hori1,2, Kazuki Hayashi3,4, Itaru Naitoh3
1Department of Gastroenterology and Metabolism, Nagoya City University Graduate School of Medical Sciences, 1 Kawasumi, Mizuho-cho, Mizuho-ku, Nagoya, 467-8601, Japan. yhori@med.nagoya-cu.ac.jp.
Background And Aims:
Migration of duodenal covered self-expandable metallic stents (C-SEMS) is the main cause of stent dysfunction in patients with malignant gastric outlet obstruction. However, the ideal method to prevent migration has not been clarified. We aimed to evaluate the feasibility and safety of duodenal C-SEMS fixation in this experimental study.
Methods:
We used the over-the-scope clip (OTSC), suture, and clip methods to fix duodenal C-SEMS and evaluated the gripping force of each device and invasion depth based on pathological findings.
Results:
The OTSC and suturing systems had a significantly higher mean gripping force compared with the clipping system (OTSC vs. clip: 13.2 vs. 1.0 Newtons [N], P < 0.001; suture vs. clip: 8.5 vs. 1.0 N, P < 0.001). OTSC compression was stronger compared with suturing (OTSC vs. suture: 13.2 vs. 8.5 N, P = 0.006). The submucosal layer, but not the muscle layer, was compressed more widely and deeply by OTSC compared with clips based on pathological findings by hematoxylin and eosin staining.
Conclusion:
Both OTSC and suturing methods used for duodenal C-SEMS fixation were feasible compared with the clipping method. The pathological evaluation of invasion depth indicated that OTSC may be safe even for preventive use. This study suggested that these methods can be applied clinically for duodenal C-SEMS fixation.
Related Concept Videos
Survey Safety
Bonding in Metals
Metallic Solids
All metallic solids exhibit high thermal and electrical conductivity, metallic luster, and malleability....
Alkali Metals
Table 1: Properties of the alkali metals
Fixation and Sectioning
The simplest type of preparation is the wet mount, in which the specimen is placed in a drop of liquid on the slide. A liquid specimen can be directly deposited on the slide using a dropper. Solid specimens, such as skin scraping, can be placed on the slide before adding a drop of liquid to prepare the wet mount. Sometimes the liquid is simply water, but stains are often added...
Metal-Ligand Bonds
In these complexes, transition metals form coordinate covalent bonds, a kind of Lewis acid-base interaction in which both of the electrons in the bond are contributed by a donor (Lewis base) to an electron acceptor (Lewis acid). The Lewis acid in...