Precision of In Vivo Quantitative Tooth Wear Measurement Using Intra-Oral Scans
Published on: July 12, 2022
Comprehensive analysis of laserscanner validity used for measurement of wear
Shu-Min Hsu1, Fan Ren2, Nader Abdulhameed1
1Department of Restorative Dental Sciences, University of Florida College of Dentistry, Gainesville, Florida.
Objectives:
The aims of this study were to test the hypotheses that (a) a laserscanner used for measuring maximum depth and volume loss will yield the same results as a surface profilometer; (b) the surface roughness will affect the maximum depth and volume loss measured with the laserscanner; (c) analytical results using the laserscanner from multiple operators have no more than 10% inter-rater difference and; (d) replicating samples using either stone or impression material is an accurate method for measuring wear using the laserscanner.
Materials And Methods:
The volume and maximum depth of indentations from fine, medium and rough burs on glass-ceramic disks were measured using two devices, a surface profilometer (Dektak II, Veeco) and a 3D Laserscanner (LAS-20, SD Mechatronik). Replicates of the indentations made from polyvinysiloxane impression material and gypsum were also measured.
Results:
Comparison of profilometer and laserscanner readings using ceramic disks demonstrated a mean error of 13.61% for depth and 25.32% for volume. Replication errors were minimal (2.6% for impression, 2.5% for stone). Surface profilometer data for volume measurements revealed a difference of 6.1% for impression and 6.5% for stone compared with ceramics. However, when measurements for replicates were compared between laserscanner and surface profilometer, depth had a mean error of 74% for impression and 51% for stone. Volume differences of 78% for impression and 44% for stone were recorded.
Conclusion:
This work demonstrated that the laserscanner was a convenient device for measuring wear but there is a need to validate the accuracy of the measurements.
More Related Videos
08:47Mimicking and Measuring Occlusal Erosive Tooth Wear with the "Rub&Roll" and Non-contact Profilometry
Published on: February 2, 2018
08:28Validated Immunochemical Assay for Comprehensive Determination of the Human Epidermal Growth Factor Receptor 2 Released from and Bound to Cells
Published on: May 9, 2025
Related Concept Videos
Reliability and Validity
Data Validation
Key parameters for method validation include:
Data Validation
Nursing assessment guides are generally based on holistic models rather than medical...
Formulating and Validating Nursing Diagnosis I
There are thirteen domains...
Formulating and Validating Nursing Diagnosis II
Risk nursing diagnoses represent clinical judgments of an individual, family, or community more vulnerable to developing the health problem than others...
In Vitro Drug Release Testing: Overview, Development and Validation