Implementation of Semiautomated Antimicrobial Susceptibility Interpretation Hardware for Nontuberculous Mycobacteria May Overestimate Susceptibility
Related Experiment Video
Updated: Jan 29, 2026

Rapid Antimicrobial Susceptibility Testing by Stimulated Raman Scattering Imaging of Deuterium Incorporation in a Single Bacterium
Published on: February 14, 2022
Mariëlle Rockland1, Mike Marvin Ruth1, Nicole Aalders1
1Department of Medical Microbiology, Radboud University Medical Center, Nijmegen, the Netherlands.
Abstract:
Nontuberculous mycobacteria (NTM) cause severe opportunistic infections and have a rising incidence in most settings. Rising diagnostic need must be met by national reference laboratories, which rely on Clinical and Laboratory Standards Institute (CLSI) guideline-approved manual readout of microtiter plates for antimicrobial susceptibility testing (AST) to determine antibiotic minimum inhibitory concentrations (MICs). Interpretation of these plates leads to different outcomes between laboratories. The SensiTitre Vizion digital MIC viewing system (Vizion) offers a more streamlined approach using semiautomated reading. Here, we conducted a blinded trial comparing the outcome of AST between manual readout and Vizion readout for 132 NTM isolates, amounting to 727 individual tests for antibiotic susceptibility ranging across 13 individual antibiotics with established CLSI breakpoints. From this, we calculated specificity, sensitivity, positive predictive value (PPV), negative predictive value (NPV) and the F1 value, as well as assessing major error (ME) and very major error (VME) rates. We find that Vizion-assisted AST produces significantly lower MICs (paired Wilcox signed rank test; P < 0.0001). The Vizion had an accuracy of 89,40%, producing 61 MEs (8.39%) and 16 VMEs (2.20%). The calculated specificity was 0.8370, the sensitivity was 0.9550, the PPV was 0.8460, the NPV was 0.9520, and the F1 score was 0.8970. We show that discrepant readings mostly stem from CLSI guideline breakpoints being close to, or overlapping, the MIC50 values, leading to small discrepancies crossing the breakpoint, contributing to VMEs and MEs. Using the Vizion in standard clinical diagnostics for NTM might lead to an overestimation of antibiotic susceptibility.
More Related Videos
Related Concept Videos
Magnetic Susceptibility and Permeability
When diamagnetic materials are placed under an external magnetic field, the moments opposite to the field are induced. Hence, the susceptibility for diamagnets has a minimal negative value of 10-5–10-6. Since...
Susceptibility, Permittivity and Dielectric Constant
Antimicrobial Effectiveness
Nursing Implementation
The five steps to implementing effective nursing care include reassessing the patient, reviewing and revising the existing nursing care plan, organizing the resources and care delivery, anticipating and preventing complications, and implementing nursing interventions.
Interpreting R Charts
An R chart plots the range of subsets of measurements collected from a process. Each point on the chart represents the range—defined as the difference between the maximum and minimum...
Antimicrobial Proteins
Interferons
Interferons (IFNs) are proteins produced by lymphocytes, macrophages, and fibroblasts infected with viruses. While IFNs cannot prevent viruses from entering and...