Combined Conventional and Amplitude-Integrated EEG Monitoring in Neonates: A Prospective Study
Sarah Grace Buttle1, Brigitte Lemyre2,3, Erick Sell1,3
11 Division of Neurology, Children's Hospital of Eastern Ontario, University of Ottawa, Ottawa, Ontario, Canada.
Background/Objective:
Seizure monitoring via amplitude-integrated EEG is standard of care in many neonatal intensive care units; however, conventional EEG is the gold standard for seizure detection. We compared the diagnostic yield of amplitude-integrated EEG interpreted at the bedside, amplitude-integrated EEG interpreted by an expert, and conventional EEG.
Methods:
Neonates requiring seizure monitoring received amplitude-integrated EEG and conventional EEG in parallel. Clinical events and amplitude-integrated EEG were interpreted at bedside. Subsequently, amplitude-integrated EEG and conventional EEG were independently analyzed by experienced neonatology and neurology readers. Sensitivity and specificity of bedside amplitude-integrated EEG as compared to expert amplitude-integrated EEG interpretation and conventional EEG were evaluated.
Results:
Thirteen neonates were monitored for an average duration of 33 hours (range 15-94, SD 25). Fourteen seizure-like events were detected by clinical observation, and 12 others by bedside amplitude-integrated EEG analysis. One of the clinical, and none of the bedside amplitude-integrated EEG events were confirmed as seizures on conventional EEG. Post hoc expert amplitude-integrated EEG interpretation revealed eight suspected seizures, all different from the ones detected by the bedside amplitude-integrated EEG team, of which one was confirmed via conventional EEG. Eight seizures were recorded on conventional EEG. Expert amplitude-integrated EEG interpretation had a sensitivity of 13% with 46% specificity for individual seizure detection, and a sensitivity of 50% with 46% specificity for detecting patients with seizures.
Conclusion:
Real-world bedside amplitude-integrated EEG monitoring failed to detect all seizures evidenced via conventional EEG, while misclassifying other events as seizures. Even post hoc expert amplitude-integrated EEG interpretation provided limited sensitivity and specificity. Considering the poor sensitivity and specificity of bedside amplitude-integrated EEG interpretation, combined monitoring may provide limited clinical benefit.
Related Concept Videos
Pulse amplitude and quality
A weak or absent pulse may indicate reduced cardiac output or poor left ventricular contraction, which can be signs of cardiovascular dysfunction or...
Sign Convention
The normal force acts perpendicular to the beam's cross-section and can...
Enolate Mechanism Conventions
Integration by Parts: Indefinite Integrals
Integration by Parts: Definite Integrals
The Integrated Rate Law: The Dependence of Concentration on Time