Automatic needle detection and real-time Bi-planar needle visualization during 3D ultrasound scanning of the liver
Muhammad Arif1, Adriaan Moelker2, Theo van Walsum1
1Department of Medical Informatics, Erasmus MC, University Medical Center Rotterdam, the Netherlands; Department of Radiology & Nuclear Medicine, Erasmus MC, University Medical Center Rotterdam, the Netherlands.
Abstract:
2D ultrasound (US) image guidance is used in minimally invasive procedures in the liver to visualize the target and the needle. Needle insertion using 2D ultrasound keeping the transducer position to view needle and reach target is challenging. Dedicated needle holders attached to the US transducer help to target in plane and at a specific angle. A drawback of this is that, the probe is fixed to the needle and cannot be rotated to assess the position of the needle in a perpendicular plane. In this study, we propose an automatic needle detection and tracking method using 3D US imaging to improve image guidance and visualization of the target in the liver with respect to the needle during these interventional procedures. The method utilizes a convolutional neural network for detection of the needle in 3D US images. In a subsequent step, the output of the convolutional neural network is used to detect needle candidates, which are fed into a final tracking step to determine the real needle position. The needle position is used to present two perpendicular cross-sectional planes of the 3D US image containing the needle in both directions. Performance of the method was evaluated in phantoms and in-vivo data by calculating the needle position distance and needle orientation angle between segmented needles and reference ground truth needles, which were manually annotated by an observer. The method successfully detects the needle position and orientation with mean errors of 1 mm and 2°, respectively. The proposed method yields a robust automatic needle detection and visualization at a frame rate of 3 Hz in 3D ultrasound imaging of the liver.
Related Concept Videos
Automatic Processing and Automatic Social Behavior
Voltammetric Techniques: Linear-Scan (E vs Time)
Real Time RT-PCR
The real-time quantification of the number of amplified products is...
Crystal Field Theory - Tetrahedral and Square Planar Complexes
Crystal field theory (CFT) is applicable to molecules in geometries other than octahedral. In octahedral complexes, the lobes of the dx2−y2 and dz2 orbitals point directly at the ligands. For tetrahedral complexes, the d orbitals remain in place, but with only four ligands located between the axes. None of the orbitals points directly at the tetrahedral ligands. However, the dx2−y2 and dz2 orbitals (along the Cartesian axes) overlap with the ligands less than the dxy,...
Leaky Scanning
Planar Rigid-Body Motion
Planar motion is typically divided into three distinct categories. The first is rectilinear translation, demonstrated by a subway train that moves along...