Conducting Respiratory Oscillometry in an Outpatient Setting
Published on: April 8, 2022
Comparison of Forced and Impulse Oscillometry Measurements: A Clinical Population and Printed Airway Model Study
Marcia Soares1, Matthew Richardson1, James Thorpe2
1NIHR Biomedical Research Centre: Respiratory Theme and Department of Respiratory Sciences, Universityof Leicester, Leicester, United Kingdom.
Abstract:
The use of commercialised forced oscillation (FOT) devices to assess impedance in obstructive diseases such as asthma has gained popularity. However, it has yet to be fully established whether resistance and reactance measurements are comparable across different FOT devices, particularly in disease. We compared two commercially available FOT devices: Impulse Oscillometry (IOS) and TremoFlo FOT (Thorasys) in a) clinical adult population of healthy controls (n = 14), asymptomatic smokers (n = 17) and individuals with asthma (n = 73) and b) a 3D printed CT-derived airway tree model resistance, as well as a 3 L standardised volume reactance. Bland-Altman Plots and linear regressions were used to evaluate bias between the devices. Resistance measurements at both 5 and 20 Hz were numerically higher with IOS compared to FOT, with evidence of small and statistically significant proportional systematic bias and a positive Bland-Altman regression slope at both 5 and 20 Hz. In contrast, the IOS device recorded reactances that were less negative at both 5 Hz and 20 Hz and significantly smaller reactance areas when compared to TremoFlo. Larger statistically significant proportional systematic biases were demonstrated with both reactance at 5 Hz and reactance area (AX) between the devices with a negative Bland-Altman regression slope. The printed airway resistance and standardised volume reactance confirmed the observations seen in patients. We have demonstrated that the impulse oscillation system and TremoFlo FOT demonstrate comparative bias, particularly when comparing airway reactance in patients. Our results highlight the need for further standardisation across FOT measurement devices, specifically using variable test loads for reactance standardisation.
More Related Videos
Related Concept Videos
Comparison Between Electrical And Gravitational Forces
Since both are inverse square law forces, the distance gets canceled when the ratio of the two forces is considered. Instead, the ratio of the electrical and gravitational forces depends on...
Impulse
Additionally, it can be shown that the...
Impulse Response
Kirchhoff's law forms an input signal equation, with the capacitor's current and voltage providing the output. Substituting the current and dividing by RC yields a differential equation. The output for an impulse input is the impulse...
What is Population Genetics?
Pharmacokinetic Models: Comparison and Selection Criterion
Physiological models take a detailed approach by considering specific molecular processes. They can predict drug distribution, metabolism, and elimination changes, providing a comprehensive understanding of how drugs interact with the body.
Intermolecular Forces