The comparative accuracy of a handheld and console ultrasound device for neuraxial depth and landmark assessment
B Carvalho1, K M Seligman2, C F Weiniger3
1Department of Anesthesiology, Perioperative, and Pain Medicine, Stanford University School of Medicine, Stanford, USA.
Background:
The study aimed to compare the accuracy of epidural depth estimation of a handheld ultrasound device, with an integrated algorithm that estimates epidural depth (AU; Accuro, Rivanna Medical), to that of a console ultrasound machine (GU; GE LOGICTM S8).
Methods:
Women requesting labor epidural analgesia consented to this prospective cohort study. The L2/3, L3/4, and L4/5 interspaces and the respective depths to the epidural space were identified, marked and measured using an AU and GU. An anesthesia provider who was blinded to ultrasound depth measurements performed epidural analgesia at one of the ultrasound identified insertion points and recorded the Tuohy needle depth at loss-of-resistance. Bland Altman analysis was used to measure the agreement between the epidural depths measured by the AU and GU.
Results:
A total of 47 women were analyzed. The mean ± standard deviation body mass index of the study cohort was 29 ± 5 kg/m2 [range 23-45]. The mean difference between the epidural depths measured by the AU and GU was -0.29 cm [95% limit of agreement 0.50 to -0.91]. The mean difference between the depth to the epidural space measured by the GU versus the needle depth was -0.33 cm [95% CI -0.49 to -0.16]. The previously reported AU versus needle depth was -0.61 cm [95% CI -0.79 to -0.44].
Conclusion:
The AU and GU provided comparable epidural depth estimates. The AU device may be a reasonable alternative to more sophisticated ultrasound devices in determining the epidural space and depth in a non-obese obstetric population.
More Related Videos
05:22Author Spotlight: Demonstrating Systematic Endobronchial Ultrasound to New Endoscopists
Published on: August 11, 2023
05:56Evaluation of the Feasibility, Safety, and Accuracy of an Intraoperative High-intensity Focused Ultrasound Device for Treating Liver Metastases
Published on: January 9, 2019
Related Concept Videos
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Uncertainty in Measurement: Accuracy and Precision
Design Example: Vintage Mixing Console
The specifications for the pre-amplifier were clear. It needed to amplify the audio signal by a factor of 10, have an input impedance above 10...
Improving Translational Accuracy
Improving Translational Accuracy
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: