Related Experiment Video
Updated: Jan 29, 2026

Reverse Total Shoulder Arthroplasty
Published on: July 5, 2011
Treatment of severe glenoid deficiencies in reverse shoulder arthroplasty: the Glenius Glenoid Reconstruction System
Philippe Debeer1, Bart Berghs2, Nicole Pouliart3
1Orthopaedics, University Hospitals Leuven, Department of Development and Regeneration, K.U. Leuven, Belgium & Institute for Orthopaedic Research and Training, Leuven, Belgium.
Background:
The treatment of glenoid bone deficiencies in primary or revision total shoulder arthroplasty is challenging. This retrospective study evaluated the short-term clinical and radiologic results of a new custom-made patient-specific glenoid implant.
Methods:
We treated 10 patients with severe glenoid deficiencies with the Glenius Glenoid Reconstruction System (Materialise NV, Leuven, Belgium). Outcome data included a patient-derived Constant-Murley score, a visual analog score (VAS), a satisfaction score, the 11-item version of the Disabilities of the Arm, Shoulder and Hand score, and the Simple Shoulder Test. We compared the postoperative position of the implant with the preoperative planned position on computed tomography scans.
Results:
At an average follow-up period of 30.5 months, the mean patient-derived Constant-Murley score was 41.3 ± 17.5 points (range, 18-76 points) with a visual analog scale of 3.3 ± 2.5 points (range, 0-7 points). The mean 11-item version of the Disabilities of the Arm, Shoulder and Hand score was 35.8 ± 18.4 (range, 2-71), and the mean Simple Shoulder Test was 47.5% ± 25.3% (range, 8%-92%). Eight patients reported the result as better (n = 3) or much better (n = 5). One patient had an elongation of the brachial plexus, and 1 patient had a period of instability. The average preoperative glenoid defect size was 9 ± 4 cm3 (range, 1-14 cm3). The mean deviation between the preoperative planned and the postoperative version and inclination was 6° ± 4° (range 1°-16°) and 4° ± 4° (range 0°-11°), respectively.
Conclusion:
Early results of the Glenius Glenoid Reconstruction System are encouraging. Adequate pain relief, a reasonable functionality, and good patient satisfaction can be obtained in these difficult cases. Further follow-up will determine the bony ingrowth and subsequent longevity of this patient-specific glenoid component.
Related Concept Videos
Muscles of the Shoulder
Anterior Thoracic Muscles
The anterior thoracic muscles include the serratus anterior, subclavius, and...
What is an Experiment?
Reversible and Irreversible Processes
Diode: Reverse bias
Thomson's e/m Experiment
A particle with charge q, speed v, and mass m enters an area from the top, where the magnetic and electric fields are perpendicular both to the particle's motion and to one another. The magnetic...
Reconstruction of Signal using Interpolation