Internal insulation breaches in an implantable cardioverter-defibrillator lead with redundant conductors
Robert G Hauser1, Jay Sengupta1, Edward J Schloss2
1Minneapolis Heart Institute Foundation, Minneapolis, Minnesota.
Background:
Internal insulation breaches (IBR) may result in implantable cardioverter-defibrillator lead failure and adverse clinical events. Concerns exist that the Durata lead may be prone to IBR.
Objective:
The goals of this study were to assess Durata failures in the Food and Drug Administration Manufacturer and User Facility Device Experience (MAUDE) database and compare them to failures in MAUDE for Endotak Reliance and Sprint Quattro Secure (QS) leads.
Methods:
We searched the MAUDE database from 2008 to 2018 for IBR and other failure modes. Included were explanted leads whose manufacturers found an insulation or conductor defect not caused by extrinsic factors.
Results:
The MAUDE search found 1011 qualifying leads. The cause of failure differed among leads (P < .001). The primary cause of Durata failure was IBR (293 of 316 leads [93%]), with IBR accounting for 47% (137 of 293); few QS (9 of 523 [1.7%]) and no Endotak Reliance leads failed because of IBR (P < .001). Durata IBR were responsible for 11 failures to treat ventricular tachycardia/ventricular fibrillation, and all were caused by high-voltage (HV) shorts between the proximal superior vena cava coil and a distal right ventricular coil cable (n = 10) or sensing conductor (n = 1); low values of HV impedance were found in these leads during defibrillation threshold testing (n = 3), after a shock or aborted shock (n = 7), and by an alert (n = 1). Inappropriate therapy was caused by 51 Durata IBR, but no QS IBR.
Conclusion:
Durata implantable cardioverter-defibrillator leads are susceptible to IBR that may result in failure to treat ventricular tachycardia/ventricular fibrillation or inappropriate therapy; such failures may occur without forewarning. HV testing of Durata leads may be indicated during pulse generator replacement or when an insulation defect is suspected.
Related Concept Videos
Conductors and Insulators
Most metals are conductors. Their atomic configuration is such that one or more electron(s) are loosely bound to the nucleus in each atom. Thus, a sea of mobile electrons are available in them, known as free electrons. Their easy...
Insulation Coordination
Charge on a Conductor
Thermal Insulation in Masonry Walls
External insulation can be applied using an Exterior Insulation and Finish System (EIFS), which involves affixing panels of plastic foam to the wall and covering them with a polymeric stucco reinforced with glass fiber mesh....
Internal Energy
Electric Field Inside a Conductor
Suppose a piece of metal is placed near a positive charge. The free electrons in the metal are attracted to the external positive charge and migrate freely toward that region. This region then...