X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
An analytical approach for the simulation of realistic low-dose fluoroscopic images
Sai Gokul Hariharan1,2, Norbert Strobel3,4, Christian Kaethner3
1Computer Aided Medical Procedures, Technische Universität München, Munich, Germany. saigokul.hariharan@tum.de.
Purpose:
The quality of X-ray images plays an important role in computer-assisted interventions. Although learning-based denoising techniques have been shown to be successful in improving the image quality, they often rely on pairs of associated low- and high-dose X-ray images that are usually not possible to acquire at different dose levels in a clinical scenario. Moreover, since data variation is an important requirement for learning-based methods, the use of phantom data alone may not be sufficient. A possibility to address this issue is a realistic simulation of low-dose images from their related high-dose counterparts.
Method:
We introduce a novel noise simulation method based on an X-ray image formation model. The method makes use of the system parameters associated with low- and high-dose X-ray image acquisitions, such as system gain and electronic noise, to preserve the image noise characteristics of low-dose images.
Results:
We have compared several corresponding regions of the associated real and simulated low-dose images-obtained from two different imaging systems-visually as well as statistically, using a two-sample Kolmogorov-Smirnov test at 5% significance. In addition to being visually similar, the hypothesis that the corresponding regions-from 80 pairs of real and simulated low-dose regions-belonging to the same distribution has been accepted in 81.43% of the cases.
Conclusion:
The results suggest that the simulated low-dose images obtained using the proposed method are almost indistinguishable from real low-dose images. Since extensive calibration procedures required in previous methods can be avoided using the proposed approach, it allows an easy adaptation to different X-ray imaging systems. This in turn leads to an increased diversity of the training data for potential learning-based methods.
More Related Videos
07:31Author Spotlight: Advancing Cell Membrane Biophysics - Exploring Interactions and Challenges Through Experimental and Computational Approaches
Published on: September 1, 2023
10:05Anatomically Realistic Neonatal Heart Model for Use in Neonatal Patient Simulators
Published on: February 5, 2019
Related Concept Videos
Development of Analytical Methods
Dose Size and Dosing Frequency: Determination Methods
Analyte Adsorption and Distribution
Jung's Analytical Theory
Determination of Multiple Dosing Parameters: Loading and Maintenance Doses
Rational Dosage Regimen: Maintenance Dose and Loading Dose
In most cases, drugs are administered repetitively or infused continuously to maintain a steady-state concentration in the body. At a steady...