Adding a protective K-wire during opening high tibial osteotomy increases lateral hinge resistance to fracture
Edouard Dessyn1,2, Akash Sharma1,2, Mathias Donnez1,2
1Department of Orthopedics and Traumatology, Institute of movement and locomotion, St Marguerite Hospital, 270 Boulevard Sainte Marguerite, BP 29, 13274, Marseille, France.
Purpose:
It was hypothesized in this in-vitro study that positioning a K-wire intersecting the cutting plane at the theoretical lateral hinge location would limit the cut depth and help preserve the lateral hinge during the opening of the osteotomy. Objectives were (1) to compare the mechanical resistance of the hinge and the protective effect of leaving the K-wire during the opening procedure (2) to check if the K-wire would limit the depth of the osteotomy.
Methods:
An ex-vivo mechanical study, testing 5 pairs of fresh-frozen tibias, was designed. CT-scan based Patient-specific cutting guides were obtained to define the cutting plane and the location of the K-wire at the hinge, using standardized 3D planning protocol. In each pair, OWHTO was performed either with or without the K-wire. To evaluate the hinge's resistance to fracture, the specimens were rigidly fixed at the proximal tibia and a direct load was applied on the free tibial diaphysis to open the osteotomy. The maximum load at breakage, maximum permissible displacement and maximal angulation of the osteotomy before hinge failure was measured. To assess the preservation of an unscathed hinge (protected by the K-wire), the distance from the end of the osteotomy cut to the lateral tibial cortical was measured in mm.
Results:
The maximum load to hinge breakage in the K-wires PsCG knees compared to the control group (48.3 N vs 5.5 N, p = 0.004), the maximum permissible displacement (19.8 mm vs 7.5 mm, p = 0.005) and the maximal angulation of the osteotomy before hinge breakage (9.9° vs 2.9°, p = 0.002) were all statistically superior in the K-wires PsCG knees compared to the control group. A mean distance of 10 ± 1 mm between cut-bone (saw-print) and lateral hinge cortical bone was found post-performing the osteotomy and the hinge failing.
Conclusion:
The maximum load to breakage and the maximum permissible displacement were, respectively, 880% and 260% higher during the opening of the OWHTO in using K-wires compared to the non-K-wire control group. This confirms the mechanical advantage of using a K-wire for both stabilization and protecting the Hinge during OWHTO. This comparative cadaveric study shows an improvement of the lateral hinges resistance to failing during the opening of the osteotomy. This can be achieved by the placement of a K-wire intersecting the cutting plane at the theoretical location of the lateral hinge.
More Related Videos
05:43Protocol for Developing a Femur Osteotomy Model in Wistar Albino Rats
Published on: August 31, 2022
05:45Author Spotlight: Exploring Regeneration in Axolotls Through Insights in Cellular and Molecular Mechanisms, Bone Healing, and Implications for Human Therapies
Published on: April 12, 2024
Related Concept Videos
Increasing Function
Lateralization
Magnetic Field Due to Two Straight Wires
Household Wiring And Electrical Safety
Magnetic Field Due To A Thin Straight Wire
Protection of Alcohols
Protection
It defines a protecting group as the masking agent to make the more reactive species inert to a given set of conditions. This concept is depicted via the illustration of liquid flow through different outlets in an assembly of pipes. The analogy helps to understand the role...