Cortical Bone Assessment Using Ultrasonic Guided Waves: A Reproducibility Study in a Healthy Population
Published on: January 31, 2025
Inter-platform reproducibility of ultrasonic attenuation and backscatter coefficients in assessing NAFLD
Aiguo Han1, Yingzhen N Zhang2, Andrew S Boehringer2
1Bioacoustics Research Laboratory, Department of Electrical and Computer Engineering, University of Illinois at Urbana-Champaign, 306 North Wright Street, Urbana, IL, 61801, USA. han51@illinois.edu.
Objectives:
To assess inter-platform reproducibility of ultrasonic attenuation coefficient (AC) and backscatter coefficient (BSC) estimates in adults with known/suspected nonalcoholic fatty liver disease (NAFLD).
Methods:
This HIPAA-compliant prospective study was approved by an institutional review board; informed consent was obtained. Participants with known/suspected NAFLD were recruited and underwent same-day liver examinations with clinical ultrasound scanner platforms from two manufacturers. Each participant was scanned by the same trained sonographer who performed multiple data acquisitions in the right liver lobe using a lateral intercostal approach. Each data acquisition recorded a B-mode image and the underlying radio frequency (RF) data. AC and BSC were calculated using the reference phantom method. Inter-platform reproducibility was evaluated for AC and log-transformed BSC (logBSC = 10log10BSC) by intraclass correlation coefficient (ICC), Pearson's correlation, Bland-Altman analysis with computation of limits of agreement (LOAs), and within-subject coefficient of variation (wCV; applicable to AC).
Results:
Sixty-four participants were enrolled. Mean AC values measured using the two platforms were 0.90 ± 0.13 and 0.94 ± 0.15 dB/cm/MHz while mean logBSC values were - 30.6 ± 5.0 and - 27.9 ± 5.6 dB, respectively. Inter-platform ICC was 0.77 for AC and 0.70 for log-transformed BSC in terms of absolute agreement. Pearson's correlation coefficient was 0.81 for AC and 0.80 for logBSC. Ninety-five percent LOAs were - 0.21 to 0.13 dB/cm/MHz for AC, and - 9.48 to 3.98 dB for logBSC. The wCV was 7% for AC.
Conclusions:
Hepatic AC and BSC are reproducible across two different ultrasound platforms in adults with known or suspected NAFLD.
Key Points:
• Ultrasonic attenuation coefficient and backscatter coefficient are reproducible between two different ultrasound platforms in adults with NAFLD. • This inter-platform reproducibility may qualify quantitative ultrasound biomarkers for generalized clinical application in patients with suspected/known NAFLD.
Related Concept Videos
Transcription Attenuation in Prokaryotes
There are several different mechanisms used to attenuate transcription. In ribosome mediated...
Coefficient of Variation
The coefficient of variation is a practical statistical tool in finance. It allows investors to assess the volatility or...
Coefficient of Correlation
If you suspect a linear relationship between x and y, then r can measure how strong the linear relationship is.
What the VALUE of r tells us:
The value of r is always between –1 and +1: –1 ≤ r ≤ 1.
The size of the correlation r indicates the...
Confidence Coefficient
Thermodynamics: Activity Coefficient
The activity coefficient is a measure of the deviation from ideal behavior. When the ionic strength of the solution is minimal, the activity coefficient of an ionic species is close to unity, making...
Factors Affecting Activity Coefficient
The activity coefficient value for an ion is close to one when the solution has almost zero ionic strength, i.e., when the solution shows close to ideal behavior. As the ionic strength of the solution increases from 0 to 0.1 mol/L, a...