Quantification in cardiovascular magnetic resonance: agreement of software from three different vendors on assessment
Leonora Zange1, Fabian Muehlberg1, Edyta Blaszczyk1,2
1Working Group on Cardiovascular Magnetic Resonance, Experimental and Clinical Research Center, a joint cooperation between the Charité Medical Faculty and the Max-Delbrueck Center for Molecular Medicine and HELIOS Hospital Berlin-Buch, Department of Cardiology and Nephrology, Medical University Berlin, Charité Campus Buch, Lindenberger Weg 80, 13125, Berlin, Germany.
Background:
Quantitative results of cardiovascular magnetic resonance (CMR) image analysis influence clinical decision making. Image analysis is performed based on dedicated software. The manufacturers provide different analysis tools whose algorithms are often unknown. The aim of this study was to evaluate the impact of software on quantification of left ventricular (LV) assessment, 2D flow measurement and T1- and T2-parametric mapping.
Methods:
Thirty-one data sets of patients who underwent a CMR Scan on 1.5 T were analyzed using three different software (Circle CVI: cvi42, Siemens Healthineers: Argus, Medis: Qmass/Qflow) by one reader blinded to former results. Cine steady state free precession short axis images were analyzed regarding LV ejection fraction (EF), end-systolic and end-diastolic volume (ESV, EDV) and LV mass. Phase-contrast magnetic resonance images were evaluated for forward stroke volume (SV) and peak velocity (Vmax). Pixel-wise generated native T1- and T2-maps were used to assess T1- and T2-time. Forty-five data sets were evaluated twice (15 per software) for intraobserver analysis. Equivalence was considered if the confidence interval of a paired assessment of two sofware was within a tolerance interval defined by ±1.96 highest standard deviation obtained by intraobserver analysis.
Results:
For each parameter, thirty data sets could be analyzed with all three software. All three software (A/B, A/C, B/C) were considered equivalent for LV EF, EDV, ESV, mass, 2D flow SV and T2-time. Differences between software were detected in flow measurement for Vmax and in parametric mapping for T1-time. For Vmax, equivalence was given between software A and C and for T1-time equivalence was given between software B and C.
Conclusion:
Software had no impact on quantitative results of LV assessment, T2-time and SV based on 2D flow. In contrast to that, Vmax and T1-time may be influenced by software. CMR reports should contain the name and version of the software applied for image analysis to avoid misinterpretation upon follow-up and research examinations.
Trial Registration:
ISRCTN12210850 . Registered 14 July 2017, retrospectively registered.
More Related Videos
11:04Quantification of Global Diastolic Function by Kinematic Modeling-based Analysis of Transmitral Flow via the Parametrized Diastolic Filling Formalism
Published on: September 1, 2014
07:55Echocardiographic Characterization of Left Ventricular Structure, Function, and Coronary Flow in Neonate Mice
Published on: April 7, 2022
Related Concept Videos
Magnetic Resonance Imaging
Assessment of the Cardiovascular System II: Inspection
Head and Neck
Assessment of the Cardiovascular System IV: Auscultation
Normal Heart Sounds
S1 (First Heart Sound)-
S1 is made by the closure of the mitral and tricuspid valves (atrioventricular valves), marking the beginning of systole.
S2 (Second Heart Sound)-
S2 is made by the closure of the aortic and pulmonic valves (semilunar valves), marking the end of the systole.
Atomic Nuclei: Magnetic Resonance
Assessment of the Cardiovascular System I: Subjective Data
Initial Enquiry
Ask the patient about their primary concern and thoroughly explore all reported symptoms.
Medical History
Investigate past illnesses affecting the cardiovascular system, such as angina, anemia, rheumatic fever, congenital heart disease, stroke, thrombophlebitis, dysrhythmias, varicosities
Inquire about symptoms...
Assessment of the Cardiovascular System III: Palpation
Jugular Venous Pressure (JVP) Measurement
Position the patient at a thirty- to forty-five-degree angle or in a semi-fowler's position. Look for the highest point of pulsation in the internal jugular vein and measure the vertical distance to the angle of Loius or sternal angle. A normal JVP is 3-4 cm above...