[Consideration on SNR in Synthetic MRI]
Akihito Ikeda1, Kohki Yoshikawa2
1Department of Radiology, Tokyo Metropolitan Health and Hospitals Corporation Ohkubo Hospital (Current address: Department of Radiology, Tokyo Metropolitan Health and Hospitals Corporation Tama-Nanbu Chiiki Hospital).
Abstract:
Synthetic magnetic resonance imaging (MRI) can create different contrast weighted images by quantifying the T1, T2, and proton density values of the subjects from a single series of scan data. It has not been clarified how the signal to noise ratio (SNR) of the synthesized image varies depending on imaging parameters. We investigated the change of SNR in synthesized MR images by the experiment using self-made phantom. The SNR ratio of synthesized image by synthetic MRI showed the same tendency as the theoretical values due to parameter change in Ny, Nx, slice thickness, number of excitations. However, as for BW, the SNR ratio tended to be different from the theoretical values in some cases. In addition, it was suggested that the SNR of the composite image has relevance to the quantitative accuracy of the T1, T2, and proton density values. We thought that this is due to the image acquisition process by synthetic MRI.
Related Concept Videos
Design Consideration
The factor of safety is another key...
Synthetic Biology
Golden rice
Golden rice is a genetically modified...
Tensile Strength Considerations of Concrete
The dimensions and shape of a concrete specimen...
Transmission Line Design Considerations
Synthetic Disvision of Polynomials
Special considerations while measuring pulse