Precision of Bone Mineral Density Measurements Around Total Ankle Replacement Using Dual Energy X-ray Absorptiometry
Carmelo Messina1, Federico Giuseppe Usuelli2, Camilla Maccario2
1IRCCS Istituto Ortopedico Galeazzi, via Riccardo Galeazzi 4, 20161 Milan, Italy; Dipartimento di Scienze Biomediche per la Salute, Università degli Studi di Milano, via Carlo Pascal, 36, 20133, Milan, Italy.
Introduction:
Joint prosthesis survival is associated with the quality of surrounding bone. Dual-energy X-ray absorptiometry (DXA) is capable to evaluate areal bone mineral density (BMD) around different prosthetic implants, but no studies evaluated periprosthetic bone around total ankle replacement (TAR). Our aim is to determine the precision of the DXA periprosthetic BMD around TAR.
Methodology:
Short-term precision was evaluated on 15 consecutive patients. Each ankle was scanned 3 times both in the posteroanterior (PA) and lateral views with a dedicated patient positioning protocol. Up to four squared regions of interest (ROIs) were placed in the periprosthetic bone around tibial and talar implants, with an additional ROI to include the calcaneal body in the lateral scan. Coefficient of variation (CV%) and least significant change were calculated according to the International Society for Clinical Densitometry.
Results:
The lateral projection showed lower mean CV values compared to the PA projection, with an average precision error of 2.21% (lateral scan) compared to 3.34% (PA scans). Overall, the lowest precision error was found at both "global" ROIs (CV = 1.25% on PA and CV = 1.3% on lateral). The highest CV value on PA was found at the medial aspect of talar side (ROI 3; CV = 4.89%), while on the lateral scan the highest CV value was found on the posterior aspect of talar side (ROI 2; CV = 2.99%).
Conclusions:
We found very good reproducibility BMD values of periprosthetic bone around TAR, that were comparable or even better compared to other studies that evaluated periprosthetic BMD around different prosthetic implants. DXA can be used to precisely monitor bone density around ankle prostheses, despite further long-term longitudinal studies are required to assess the clinical utility of such measurements.
Related Concept Videos
Strain-Energy Density
In the elastic region of a material, the relationship between the stress and the strain is linear and follows Hooke's Law. The strain energy density in this region...
Uncertainty in Measurement: Accuracy and Precision
Essential Minerals for Bone Health
Calcium and Phosphorus
Calcium is a critical component of bones, especially in the form of calcium phosphate and calcium carbonate. Since the body cannot make calcium, it must be obtained from the diet. However, calcium cannot be absorbed from the small intestine without...
Ankle Joint
X-ray Crystallography
Diffraction
Diffraction is the change in the direction of travel experienced by an electromagnetic wave when it encounters a physical barrier whose dimensions are comparable to those of the wavelength of the light. X-rays are electromagnetic radiation with wavelengths about as long as the distance between neighboring...
Water and Mineral Acquisition