Association Between the Quantity of Subcutaneous Fat and the Inter-Device Agreement of 2 Tissue Oximeters
Alexis Cournoyer1, Sylvie Cossette2, Jean Paquet3
1Université de Montréal, Montréal, Québec, Canada; Hôpital du Sacré-Cœur de Montréal, Montréal, Québec, Canada; Institut de Cardiologie de Montréal, Montréal, Québec, Canada; Hôpital Maisonneuve-Rosemont, Montréal, Québec, Canada.
Objective:
This study aimed to evaluate the association between the quantity of subcutaneous fat (assessed by skinfold thickness) and the inter-device agreement of 2 tissue oximeters.
Design:
This is a prospective cohort study.
Setting:
This study was conducted in a tertiary care academic urban hospital.
Participants:
Healthy volunteers were recruited.
Interventions:
All patients recruited had their tissue saturations and skinfold thickness measured at 4 different sites (shoulder, forearm, knee, and calf) on both sides of the body using 2 tissue oximeters, the INVOS 5100C and the EQUANOX 7600.
Measurements And Main Results:
Higher skinfold measures were associated with an increase in the difference between measures provided by both oximeters (slope = -0.59, Pearson correlation coefficient = -0.51, p < 0.001). This observed association persisted in a linear mixed model (-0.48 [95% confidence interval [CI] -0.61 to -0.36], p < 0.001). The sex of the volunteers also influenced the inter-oximeter agreement (women: -5.77 [95% CI -8.43 to -3.11], p < 0.001), as well as the forearm sites (left forearm: -7.16 [95% CI -9.85 to -4.47], p < 0.001; right forearm:-7.01 [95% CI -9.61 to -4.40], p < 0.001).
Conclusion:
The inter-device agreement of the 2 studied oximeters is correlated to the quantity of subcutaneous fat. Monitoring using tissue oximetry should be interpreted with great care when sensors are placed on sites with a significant quantity of subcutaneous fat. In addition to the monitoring of cerebral oximetry, following the variations of saturations at the same peripheral site seems to remain the most secure way to use that technology for the monitoring of critically ill patients.
Related Concept Videos
Base Quantities and Derived Quantities
The International Organization for...
Estimation of the Physical Quantities
Association Areas of the Cortex
Prefrontal Association Area: This area is located in the frontal lobe and is involved in planning, decision-making, and moderating social behavior. It connects with primary motor areas,...
Associative Learning
Classical conditioning, also known...
Relating Angular And Linear Quantities - I
When comparing the linear and rotational variables individually, the linear variable of position has physical units of meters, whereas the angular position variable has dimensionless units of radians, as it is the ratio of two lengths. The linear velocity...
Relating Angular And Linear Quantities - II