Related Experiment Video
Updated: Jan 28, 2026

Combined Supine and Standing Imaging for Varicocele: An Improved Diagnostic Approach
Published on: November 22, 2024
Reproducibility of pelvic sagittal inclination while acquiring radiographs in supine and standing postures
Keisuke Uemura1, Masaki Takao1, Yoshito Otake2
11 Department of Orthopaedic Medical Engineering, Graduate School of Medicine, Osaka University, Suita, Osaka, Japan.
Purpose:
Pelvic position on the sagittal plane is usually evaluated with the pelvic sagittal inclination (PSI) angle from a single radiograph. However, the reproducibility of pelvic positioning has not been investigated, and thus, the validity of measuring the PSI from a single film/time point is not understood. Herein, the reproducibility of a patient's pelvic positions in supine and standing postures was analyzed.
Methods:
A total of 34 patients who underwent either a pelvic osteotomy or total hip arthroplasty were enrolled in this study. Preoperative radiographs in both supine and standing postures were acquired twice (first X-ray and second X-ray) within 6 months; preoperative computed tomography (CT) images of the full pelvis were also acquired in a supine posture (preop-CT). To eliminate measurement variability, each PSI was automatically measured from radiographs and CT images through the use of CT segmentation and landmark localization followed by intensity-based 2D-3D registration. The absolute difference of PSI among each image was calculated and the intra-class correlation coefficient (ICC) in each posture was also analyzed.
Results:
The median absolute differences of PSI in the supine posture were 1.3° between the first and second X-rays, 1.2° between the first X-ray and preop-CT, and 1.3° between the second X-ray and preop-CT. The median absolute difference of PSI in the standing posture was 1.5°. The ICC was 0.965 (95% CI: 0.939-0.981) in supine and 0.977 (95% CI: 0.954-0.988) during standing.
Conclusions:
Pelvic positions in supine and standing postures are reproducible. Thus, measuring the PSI from a single radiograph is reliable.
Related Concept Videos
Inclination of a Line
Work Done Over an Inclined Plane
The work done by gravity to move a rigid body, or the work done by an opposing force to move a rigid body against gravity, can be calculated using the center-of-mass framework. It is the line integral of the force of gravity over the path, considered positive if...
Standing Waves
Moments of Inertia for an Area about Inclined Axes
Modes of Standing Waves - I
Modes of Standing Waves: II
For a tube open at one end and closed at the other filled with air, the modes are such that there is always an antinode at the open end and a node at the closed end....