Radiation Exposure and Operation Time in Percutaneous Endoscopic Lumbar Discectomy Using Fluoroscopy-Based Navigation
Hao Qin1, Shengbin Huang2, Lin Xu2
1Guangxi Medical University, Nanning, Guangxi, China; Orthopaedic Department, Guigang People's Hospital, Guigang, Guangxi, China.
Objective:
This study evaluated radiation exposure and operation time of percutaneous endoscopic lumbar discectomy (PELD) by using a fluoroscopy-based navigation system for access and localization.
Methods:
Eighty-six PELDs performed by a single surgeon were retrospectively analyzed. Patients were separated into 2 groups: group A (using a three-dimensional [3D]-printed navigation instrument and fluoroscopy-based navigation system) and group B (with conventional fluoroscopy and standard instrumentation). The operation, fluoroscopy, and total access time were collected, as well as fluoroscopy and access times.
Results:
The operative time for group A was 59 minutes (standard deviation [SD], 6 minutes) and 106 minutes (SD, 15 minutes) in group B (P < 0.001). In group A, fluoroscopy was used an average of 5 times (SD, 0.7) and 29 times (SD, 8) in group B (P < 0.001). The fluoroscopy time was 9 minutes (SD, 2 minutes) in group A and 40 minutes (SD, 8 minutes) in group B (P < 0.001). The number of access attempts was 1.3 (SD, 0.5) in group A and 8 (SD, 2 times) in group B (P < 0.001). The total access time was 11 minutes (SD, 2 minutes) in group A and 28 minutes (SD, 5 minutes) in group B (P < 0.001).
Conclusions:
PELD using the fluoroscopy-based navigation system showed lower operative, fluoroscopy, and access time compared with conventional techniques. In addition, fewer fluoroscopy images and access attempts were made in the navigation group. These data suggest that this novel technique reduces fluoroscopy and operation time and may reduce risks of repeated surgical access attempts.