Semiquantitative, fully automated urine test strip analysis
Related Experiment Video
Updated: Jan 28, 2026

Cell-Free DNA Integrity Analysis in Urine Samples
Published on: January 5, 2017
Matthijs Oyaert1, Joris R Delanghe1
1Department of Laboratory Medicine, Ghent University Hospital, Gent, Belgium.
Objectives:
Urinalysis is one of the most frequently ordered diagnostic laboratory tests. In order to reduce workload and costs, rapid screening tests such as urine test strip analyses are applied. The aim of this study was to evaluate the analytical performance of the UC-3500 as well as the diagnostic performance in comparison with reference methods.
Design And Methods:
We measured within-run and between-run imprecision based on quantitative reflectance values. 347 prospectively included urine specimens were investigated for the presence of glucose, protein, albumin, leukocyte esterase, and hemoglobin peroxidase activity, and ordinal scale results were compared to an automated urine particle analyzer (UF-5000, Sysmex, Kobe, Japan) and wet chemistry (Roche Cobas 8000, Mannheim, Germany).
Results:
Within-run and between-run imprecision results based on reflectance data for both the 9 and 11 parameter test strips ranged from 0.07% to 1.36% for the low-level control and from 0.37% to 6.13% for the high-level control, depending on the parameter. Regarding diagnostic performance, the sensitivity/specificity for glucose, protein, albumin, leukocyte esterase, and hemoglobin peroxidase was 100/60%, 94.2/88.2%, 81.8/89.2%, 81.7/92.8%, and 85.1/88.6%, respectively; the negative predictive value was 100%, 83.3%, 89.1%, 94.6%, and 96.1%. The Spearman correlation coefficients of the UC-3500 vs reference methods ranged from 0.915 to 0.967, depending on the parameter.
Conclusion:
This fully automated urine test strip analyzer overall shows a satisfying performance and can reliably screen out negative urine samples in order to focus on further characterization of positive samples in the following steps of the workflow.
Related Concept Videos
Filtration and Urine Formation
Urine Studies II: Urine Culture and Sensitivity Test
Voltammetry: Stripping Methods
Anodic Stripping Voltammetry (ASV)
ASV is used to determine metals and metalloids at trace levels. It involves two steps: deposition and stripping. First, a negative potential is applied to the...
Physiology of Urine Formation
Glomerular Filtration
The first stage in urine formation is glomerular filtration. Each kidney contains approximately 1 million nephrons, the functional units of filtration, with a...
Formation of Concentrated Urine
Urine Studies I: Urinalysis