COMPARISON BETWEEN SEVERAL OPTICAL COHERENCE TOMOGRAPHY ANGIOGRAPHY DEVICES AND INDOCYANINE GREEN ANGIOGRAPHY OF
Federico Corvi1, Mariano Cozzi, Erika Barbolini
1Department of Biomedical and Clinical Science "Luigi Sacco", Eye Clinic, Sacco Hospital, University of Milan, Milan, Italy.
Purpose:
To compare indocyanine green angiography and four different optical coherence tomography angiography (OCTA) devices and to test their reproducibility in the evaluation of choroidal neovascularization (CNV).
Methods:
This study was an observational case series of Type 1 and Type 2 CNV presenting at the Eye Clinic, Department of Biomedical and Clinical Sciences, Luigi Sacco Hospital, University of Milan, imaged with indocyanine green angiography and four different OCTA devices: prototype PlexElite (Carl Zeiss Meditec), prototype Spectralis OCTA (Spectralis; Heidelberg Engineering, Heidelberg, Germany), Optovue RTVue XR Avanti (Optovue, Inc, Fremont, CA), and AngioPlex (Cirrus 5000 HD-OCT; Carl Zeiss Meditec, Inc, Dublin, CA). Optical coherence tomography angiography examinations were performed using a 3 × 3-mm or 6 × 6-mm volume scan pattern to capture the full lesion. The mean area of CNV, vessel density, and fractal dimension were compared between the devices.
Results:
Seventeen eyes of 17 patients with CNV were enrolled. The CNV area was significantly larger in indocyanine green angiography than in the OCTA devices. Moreover, the four different instruments produced measurements with different mean values of CNV area, vessel density, and fractal dimension. The Bland-Altman analysis revealed that the limits of agreement for the comparisons were not acceptable.
Conclusion:
The choroidal neovascularization area is larger when imaged with standard indocyanine green angiography than with different OCTA devices with different algorithms, wavelengths, and scan patterns. The differences between several OCTA devices in the evaluation of quantitative parameters of CNV, such as CNV area, vessel density, and fractal dimension, suggest the necessity to standardize OCTA measurements for research and clinical practice.
More Related Videos
10:24Detecting Abnormalities in Choroidal Vasculature in a Mouse Model of Age-related Macular Degeneration by Time-course Indocyanine Green Angiography
Published on: February 19, 2014
09:56In Vivo Multimodal Imaging and Analysis of Mouse Laser-Induced Choroidal Neovascularization Model
Published on: January 21, 2018
Related Concept Videos
Green Algae
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Evolutionary Relationships through Genome Comparisons
Comparison Between Electrical And Gravitational Forces
Since both are inverse square law forces, the distance gets canceled when the ratio of the two forces is considered. Instead, the ratio of the electrical and gravitational forces depends on...
Comparison between RL and RC circuits