Ultrasonic cartilage thickness measurement is accurate, reproducible, and reliable-validation study using
Simon Damian Steppacher1, Markus Simon Hanke2, Corinne Andrea Zurmühle2
1Department of Orthopaedic Surgery and Traumatology, Inselspital, Bern University Hospital, University of Bern, CH-3010, Bern, Switzerland. simon.steppacher@insel.ch.
Background:
Ultrasonography is a fast and patient-friendly modality to assess cartilage thickness. However, inconsistent results regarding accuracy have been reported. Therefore, we asked what are (1) the accuracy, (2) reproducibility, and (3) reliability of ultrasonographic cartilage thickness measurement using contrast-enhanced micro-CT for validation?
Methods:
A series of 50 cartilage-bone plugs were harvested from fresh bovine and porcine joints. Ultrasonic cartilage thickness was determined using an A-mode, 20-MHz hand-held ultrasonic probe with native (1580 m/s) and adjusted speed of sound (1696 m/s). All measurements were performed by two observers at two different occasions. Angle of insonation was controlled by tilting the device and recording minimal thickness. Retrieval of exact location for measurement was facilitated by aligning the circular design of both cartilage-bone plug and ultrasonic device. There was no soft tissue interference between cartilage surface and ultrasonic probe. Ground truth measurement was performed using micro-CT with iodine contrast agent and a voxel size of 16 μm. The mean cartilage thickness was 1.383 ± 0.402 mm (range, 0.588-2.460 mm).
Results:
Mean accuracy was 0.074 ± 0.061 mm (0.002-0.256 mm) for native and 0.093 ± 0.098 mm (0.000-0.401 mm) for adjusted speed of sound. Bland-Altman analysis showed no systematic error. High correlation was found for native and adjusted speed of sound with contrast-enhanced micro-CT (both r = 0.973; p < 0.001). A perfect agreement for reproducibility (intraclass correlation coefficient [ICC] 0.992 and 0.994) and reliability (ICC 0.993, 95% confidence interval 0.990-0.995) was found.
Conclusions:
Ultrasonic cartilage thickness measurement could be shown to be highly accurate, reliable, and reproducible. The A-mode ultrasonic cartilage thickness measurement is a fast and patient-friendly modality which can detect early joint degeneration and facilitate decision making in joint preserving surgery.
Related Concept Videos
Reliability and Validity
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Self-Evaluation: Self-Enhancement and Self-Verification
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...