Intraoperative optical pachymetry in photorefractive keratectomy
Mark D Mifflin1, Xavier M Mortensen1
1Department of Ophthalmology and Visual Sciences, John A. Moran Eye Center, University of Utah, Salt Lake City, USA.
Purpose:
To study the effect of variables on the accuracy and reliability of the optical pachymeter built into the WaveLight EX500 excimer laser during photorefractive keratectomy (PRK).
Setting:
John A. Moran Eye Center, University of Utah, Salt Lake City, USA.
Design:
Retrospective case series.
Methods:
A chart review of 352 eyes (181 patients) that had excimer laser PRK was performed. Programmed excimer laser residual stromal bed (RSB) measurements, optical pachymeter measurements after ablation, and Scheimpflug pachymetry measurements (Pentacam) at the 1-year follow-up were compared. Variables included ablation time, preoperative spherical equivalent (SE), 1-year SE, mitomycin-C use, operating room temperature and humidity, and programmed monovision.
Results:
The mean programmed RSB was 27 μm greater than the optical pachymetry post-ablation measurement (P < .001). Of patients with a 1-year follow-up, the 1-year Scheimpflug pachymetry RSB was 24 μm greater than the optical pachymetry post-ablation RSB (P < .001). Comparison of the programmed RSB with the optical pachymetry post-ablation RSB showed that the preoperative SE and ablation time had a Pearson correlation coefficient of -0.36 and 0.30, respectively (P < .001). There was no correlation between operating room temperature, humidity, or programmed monovision with these differences.
Conclusions:
The RSB post-ablation values measured by optical pachymetry during PRK were significantly lower than the programmed excimer laser RSB value and 1-year Scheimpflug pachymetry RSB value. Intraoperative pachymetry during PRK underpredicted the actual long-term RSB thickness. The greater temporary drying effect associated with increased ablation time in higher myopic corrections might have caused this error.
Related Concept Videos
Properties of Enantiomers and Optical Activity
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...
Stereoisomerism
Isomers are different chemical species that have the same chemical formula.
Transition metal complexes often exist as geometric isomers, in which the same atoms are connected through the same types of bonds but with differences in their orientation in space. Coordination complexes with two different ligands in the cis and trans positions from a ligand of interest form isomers. For example, the octahedral [Co(NH3)4Cl2]+ ion has two isomers (Figure 1) In the cis...
Vision