Background:
Hemoglobin A1c, (HbA1c) which is the major constituent of glycated hemoglobin, has been used in the follow-up of retrospective glycemia for years and in the diagnosis of diabetes mellitus nowadays. Since the analytical performance of HbA1c should be high likewise all laboratory tests, various quality control measures are used. Sigma metrics is one of these measures and it is the combination of bias, precision and total allowable error that ensures a general evaluation of analytical quality. The aim of our study was to evaluate the analytical performance of Bio-Rad's Variant Turbo II HbA1c analyzer according to sigma metrics.
Methods:
Sigma levels were calculated using the data obtained from two levels of internal and 12 external quality control materials (Bio-Rad) of Variant II Turbo HbA1c analyzer according to σ = (TEa% - Bias%) / CV% formula.
Results:
The mean sigma levels for low and high quality control materials were found to be 3.0 and 4.1, respectively.
Conclusions:
The annual mean analytical performance of Variant II Turbo HbA1c analyzer was found to be acceptable according to sigma metrics. In order to be sure of the difference in HbA1c results indicating the success or failure in treatment but not arise from analytical variation, it is thought that more stringent quality control measures should be applied to reach higher sigma levels.
More Related Videos
10:50Purification and Analytics of a Monoclonal Antibody from Chinese Hamster Ovary Cells Using an Automated Microbioreactor System
Published on: May 1, 2019
06:02Evaluating Usability Aspects of a Mixed Reality Solution for Immersive Analytics in Industry 4.0 Scenarios
Published on: October 6, 2020
Related Concept Videos
Histone Variants at the Centromere
Development of Analytical Methods
Analyte Adsorption and Distribution
Jung's Analytical Theory
Graphical and Analytic Representation of Sinusoids
The first step is measuring the peak-to-peak value, which is twice the amplitude of the sinusoid. This provides information about the maximum voltage swing of the waveform.
Secondly, the period and angular frequency are determined. The period is the time taken for one complete cycle of the waveform, while...
Exercise and Muscle Performance
Endurance exercises
Endurance exercises involve running, swimming, or cycling, which require repetitive movements with low force output. When a person engages in endurance exercise, a few noticeable changes occur in their skeletal muscles. For instance, the number of capillaries...