Dependence of neutrons generated by 7Li(p,n) reaction on Li thickness under free-air condition in accelerator-based
Satoshi Nakamura1, Hiroshi Igaki2, Hiroyuki Okamoto1
1Department of Medical Physics, National Cancer Center Hospital, Tsukiji 5-1-1, Chuo-ku, Tokyo 104-0045, Japan; Division of Research and Development for Boron Neutron Capture Therapy, National Cancer Center Exploratory Oncology Research & Clinical Trial Center, Tsukiji 5-1-1, Chuo-ku, Tokyo 104-0045, Japan.
Purpose:
An accelerator-based boron neutron capture therapy (BNCT) system with a solid-state Li target is reported to have degradation of the Li target. The degradation reduces the Li thickness, which may change spectra of the generated neutrons corresponding to the Li thickness. This study aims to examine the relationship between the Li thickness and the generated neutrons and to investigate the effects of the Li thickness on the absorbed dose in BNCT.
Method:
The neutron energy spectra were calculated via Monte Carlo simulation for Li thicknesses ranging from 20 to 150 μm. Using the system, the saturated radioactivity of gold induced by reactions between 197Au and the generated neutrons was evaluated with the simulation and the measurement, and those were compared. Additionally, for each Li thickness, the saturated radioactivity was compared with the number of generated neutrons. The absorbed doses delivered by 10B(n,α)7Li, 14N(n,p)14C, 1H(n, g)2H, and (n,n') reactions in water were also calculated for each Li thickness.
Results:
The measurement and simulation indicated a reduction in the number of neutrons due to the degradation of the Li target. However, the absorbed doses were comparable for each Li thickness when the requisite number of neutrons for BNCT was delivered. Additionally, the saturated radioactivity of 198Au could be a surrogate for the number of neutrons even if the Li thickness was varied.
Conclusions:
No notable effect to the absorbed dose was observed when required neutron fluence was delivered in the BNCT even if the degradation of the Li was observed.
Related Concept Videos
Temperature Dependence on Reaction Rate
Atoms, molecules, or ions must collide before they can react with each other. Atoms must be close together to form chemical bonds. This premise is the basis for a theory that explains many observations regarding chemical kinetics, including factors affecting reaction rates.
The collision theory is based on the postulates that (i) the reaction rate is proportional to the rate of reactant collisions, (ii) the reacting species collide in an orientation allowing contact between...
Nuclear Power
Nuclear Fuels
Nuclear fuel consists of a fissile isotope, such as uranium-235, which must be present in sufficient quantity to provide a...
Targeted Cancer Therapies
There are several types of targeted therapies against...
Acids, Bases and Neutralization Reactions
Atomic Mass
Nuclear Transmutation