Dose calculation deviations induced by fractional image-guided-couch-shifts for Varian Halcyon MV cone beam CT
Yuliang Huang1, Hao Wu1, Qiaoqiao Hu1
1Key Laboratory of Carcinogenesis and Translational Research (Ministry of Education/Beijing), Department of Radiation Oncology, Peking University Cancer Hospital & Institute, Beijing 100142, China.
Purpose:
To assess the criticality of calculation deviations induced by fractional image-guided-couch-shifts for Halcyon MV cone beam CT (CBCT) dose, which is incorporated as part of total treatment dose.
Methods:
Eclipse-calculated imaging dose was first validated in 'Cheese Phantom' by measurement. Then, the actual imaging dose (Dact) for 18 historical patients of various sites were recalculated based on 513 MV CBCT-guided-couch-shift data, and compared with reference computations based on treatment isocentre (Dref). Patient- and plan-specific dose from treatment fields was integrated with Dact and Dref respectively for comparison.
Results:
The average absolute relative disagreements between the measured and calculated dose were less than 1.23%. The mean ± 1SD of gamma passing rates of the accumulated imaging dose and total dose were 80.71 ± 6.22% and 99.81 ± 0.32% respectively based on 3 mm/3%/local/10% threshold criteria. The accumulated errors of minimum imaging dose to PTV were no larger than -14.38 cGy, which were reduced to -0.82 cGy after the heterogeneous treatment dose was overlaid. The mean relative discrepancies of PTV minimum dose were -0.61 cGy (-0.71%) and -0.00 (0.00%), before and after incorporating the treatment dose respectively.
Conclusions:
The Eclipse-calculated Halcyon MV CBCT dose was validated. Although the isocentre displacement-induced imaging dose calculation errors for Halcyon MV CBCT were partially cancelled out by couch shifts of various directions and distances, especially after the incorporation of heterogeneous treatment dose, it was still advisable to monitor the accumulated deviations and replan when unacceptable target under-dose or organ over-dose were observed.
More Related Videos
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
11:38Voluntary Breath-hold Technique for Reducing Heart Dose in Left Breast Radiotherapy
Published on: July 3, 2014
Related Concept Videos
Calculating Standard Deviation
The standard deviation value is small when all the data is concentrated close to the mean. Here the data exhibits low variation. The standard deviation value is larger when the data values are more spread out from the mean. Here, the data displays high...
Standard Deviation of Calculated Results
A broad Gaussian distribution curve has a wider standard deviation, representing a data set with...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Variation: Normal Distribution, Range, and Standard Deviation
Imaging Studies for Cardiovascular System V: CT
Standard Deviation