Reducing distortions in echo-planar breast imaging at ultrahigh field with high-resolution off-resonance maps
Michael J van Rijssel1, Frank Zijlstra1, Peter R Seevinck1
1Center for Image Sciences, UMC Utrecht, Utrecht, The Netherlands.
Purpose:
DWI is a promising modality in breast MRI, but its clinical acceptance is slow. Analysis of DWI is hampered by geometric distortion artifacts, which are caused by off-resonant spins in combination with the low phase-encoding bandwidth of the EPI sequence used. Existing correction methods assume smooth off-resonance fields, which we show to be invalid in the human breast, where high discontinuities arise at tissue interfaces.
Methods:
We developed a distortion correction method that incorporates high-resolution off-resonance maps to better solve for severe distortions at tissue interfaces. The method was evaluated quantitatively both ex vivo in a porcine tissue phantom and in vivo in 5 healthy volunteers. The added value of high-resolution off-resonance maps was tested using a Wilcoxon signed rank test comparing the quantitative results obtained with a low-resolution off-resonance map with those obtained with a high-resolution map.
Results:
Distortion correction using low-resolution off-resonance maps corrected most of the distortions, as expected. Still, all quantitative comparison metrics showed increased conformity between the corrected EPI images and a high-bandwidth reference scan for both the ex vivo and in vivo experiments. All metrics showed a significant improvement when a high-resolution off-resonance map was used (P < 0.05), in particular at tissue boundaries.
Conclusion:
The use of off-resonance maps of a resolution higher than EPI scans significantly improves upon existing distortion correction techniques, specifically by superior correction at glandular tissue boundaries.
Related Concept Videos
Crystal Field Theory - Tetrahedral and Square Planar Complexes
Crystal field theory (CFT) is applicable to molecules in geometries other than octahedral. In octahedral complexes, the lobes of the dx2−y2 and dz2 orbitals point directly at the ligands. For tetrahedral complexes, the d orbitals remain in place, but with only four ligands located between the axes. None of the orbitals points directly at the tetrahedral ligands. However, the dx2−y2 and dz2 orbitals (along the Cartesian axes) overlap with the ligands less than the dxy,...
Echo
Imagine the sound is reflected back to the ears. Assuming that the source is very close to the human, the difference between hearing the two sounds—the emitted sound and the reflected sound—may be more than the minimum time for perceiving distinct sounds. If this is the case,...
Resonance
Magnetic Resonance Imaging
Imaging Studies IV: Magnetic Resonance Imaging
¹H NMR: Interpreting Distorted and Overlapping Signals
As Δν decreases and the signals move closer, the doublets appear increasingly distorted. The intensities of the inner lines increase at the cost of those of the outer lines as the signals are...