Linac-based radiosurgery for multiple brain metastases: Comparison between two mono-isocenter techniques with
Ruggero Ruggieri1, Stefania Naccarato1, Rosario Mazzola1
1Department of Radiation Oncology, IRCCS 'Sacro Cuore don Calabria', Negrar (VR), Italy.
HyperArc (HA) and Multiple Brain Mets (MBM) stereotactic radiosurgery techniques offer comparable plan quality for brain metastases. HA demonstrated superior conformity, while both methods showed satisfactory dosimetric consistency in delivering radiation doses.
Area of Science:
- Radiation Oncology
- Medical Physics
- Neurosurgery
Background:
- Stereotactic radiosurgery (SRS) for multiple brain metastases (BM) utilizes mono-isocenter techniques.
- Clinically available techniques include HyperArc (HA), Multiple Brain Mets (MBM), and Monaco-HDRS.
Purpose of the Study:
- To compare the plan quality and dosimetric consistency of HyperArc (HA) and Multiple Brain Mets (MBM) SRS techniques.
- Evaluate the performance of two mono-isocenter volumetric modulated arc therapy (VMAT) techniques for treating multiple brain metastases.
Main Methods:
- Generated HA and MBM plans for 20 patients with multiple BM (2-10 lesions).
- Assessed plan quality using Paddick conformity (CI) and gradient (GI) indices, and dose to healthy brain (mean dose, V12).
- Verified dosimetric consistency using radiochromic films and gamma analysis (γ-analysis).
Main Results:
- HyperArc (HA) plans showed significantly improved conformity (CI) compared to MBM (0.94 vs. 0.75).
- No significant differences were found in gradient index (GI), mean dose, or V12 to healthy brain between HA and MBM.
- Gamma analysis revealed significantly improved passing rates for HA, indicating better dosimetric consistency.
Conclusions:
- HyperArc (HA) provides superior target conformity for SRS of multiple brain metastases.
- Both HA and MBM techniques demonstrate satisfactory dosimetric consistency between planned and delivered doses.
- The study suggests near-equal overall plan quality between HA and MBM SRS techniques, with HA offering improved conformity.
Related Concept Videos
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Multiple Allele Traits
Multiple Regression
Farmers can use multiple regression to determine the crop yield based on more than one factor, such as water availability, fertilizer, soil properties, etc. Here, the crop yield is the response or dependent variable as it depends on the other independent variables. The analysis requires the construction of a scatter plot...
Deformation of Member under Multiple Loadings
In the case of a member with a variable cross-section, the strain is not constant but depends on the position. The deformation of an...
Multiple Voltage Sources
In series, the positive terminal of one battery is connected to the negative terminal of another battery. Hence, the voltage of each battery is added to give the net voltage, which is increased because each battery boosts the electrons that enter it. The same current flows through each battery because they are connected in series.
Batteries are...
Multiple Bar Graph
Each bar or column in the multiple bar graph represents a data value. These graphs are used primarily in interrelating two or more sets of data. The categories of different kinds of data are listed along the horizontal or x-axis, whereas...