Related Experiment Video
Updated: Jan 28, 2026

Echocardiographic Evaluation of Atrial Communications before Transcatheter Closure
Published on: February 8, 2022
Left atrial appendage closure with the Watchman device using intracardiac vs transesophageal echocardiography:
Majd E Hemam1, Kenji Kuroki2, Paul A Schurmann1
1Division of Cardiac Electrophysiology, Houston Methodist DeBakey Heart and Vascular Center, Houston, Texas.
Background:
Imaging guidance for left atrial appendage (LAA) closure (LAAC) conventionally consists of transesophageal echocardiography (TEE) and fluoroscopy under general anesthesia (GA). Intracardiac echocardiography (ICE) can eliminate the need for GA, expedite procedural logistics, and reduce the patient experience to a simple venous puncture.
Objective:
The purpose of this study was to define optimal ICE views and compare procedural parameters and cost of ICE vs TEE during LAAC with the Watchman device.
Methods:
Optimal ICE views of the LAA for Watchman implant were delineated using Carto-Sound and 3-dimensional rendition of the LAA in 6 patients. Procedural and financial parameters of 104 consecutive patients with standard indications for LAAC undergoing Watchman implant using ICE guidance through a single transseptal puncture (n = 53 [51%]) were compared with those of TEE-guided implants (n = 51 [49%]) in 3 centers.
Results:
Clinical characteristics were similar between the 2 groups. Total in-room, turnaround, and fluoroscopy times all were shorter using ICE (P <.05) under local anesthesia compared to the TEE group. Implant success was 100% in both groups without peri-device leaks or procedural complications. Follow-up TEE showed no significant peri-device leak in both groups. Total hospital charges for ICE with local anesthesia vs TEE were similar, as were total hospital direct and indirect costs. Professional fees were significantly lower with ICE and local anesthesia than with TEE because the charge of anesthesia staff was avoided.
Conclusion:
ICE-guided Watchman implant is safe, feasible, and comparable in cost to TEE during LAAC with a Watchman device but avoids GA and expedites procedure turnaround.
More Related Videos
Related Concept Videos
Surface Appendages of Archaea
Design Consideration
The factor of safety is another key...
Tensile Strength Considerations of Concrete
The dimensions and shape of a concrete specimen...
Transmission Line Design Considerations
Angle Closure Glaucoma: Treatment
Special considerations while measuring pulse