Studying Cavitation Enhanced Therapy
Published on: April 9, 2021
Passive cavitation mapping using dual apodization with cross-correlation in ultrasound therapy monitoring
Shukuan Lu1, Xianbo Yu1, Renyan Li1
1The Key Laboratory of Biomedical Information Engineering of Ministry of Education, Department of Biomedical Engineering, School of Life Science and Technology, Xi'an Jiaotong University, Xi'an 710049, PR China.
Abstract:
Recently, passive acoustic mapping (PAM) has been successfully applied for dynamic monitoring of ultrasound therapy by beamforming acoustic emissions of cavitation activity during ultrasound exposure. The most widely used PAM algorithm in the literature is time exposure acoustics (TEA), which is a standard delay, sum, and integrate algorithm. However, it results in large point spread function (PSF) and serious imaging artifacts for the case where a narrow-aperture receiving array such as a standard B-mode linear array is used, therefore degrading the quality of cavitation image. To address these challenges, in this paper, we proposed a novel PAM algorithm namely dual apodization with cross-correlation (DAX)-based TEA, in which DAX was originally used as a reconstruction algorithm in medical ultrasound imaging. In the proposed algorithm, two sets of signals were beamformed by two receive apodization functions with alternating elements enabled, and the cross-correlation coefficient of the two signals served as a weighting factor that would be multiplied to the sum of the two signals. The performance of the proposed algorithm was tested on simulated channel data obtained using a multi-bubble model, and experiments were also performed in an in vitro vessel phantom with flowing microbubbles as cavitation nuclei. The reconstructed cavitation images were evaluated quantitatively using established quality metrics including full width at half maximum (FWHM), A-6dB area, and signal-to-noise ratio (SNR). The results suggested that the proposed algorithm significantly outperformed the conventionally used TEA algorithm. This work may have the potential of providing a useful tool for highly accurate localization of cavitation activity during ultrasound therapy.
More Related Videos
08:19Controllable Nucleation of Cavitation from Plasmonic Gold Nanoparticles for Enhancing High Intensity Focused Ultrasound Applications
Published on: October 5, 2018
07:38Real-time Monitoring of High Intensity Focused Ultrasound HIFU Ablation of In Vitro Canine Livers Using Harmonic Motion Imaging for Focused Ultrasound HMIFU
Published on: November 3, 2015
Related Concept Videos
Correlations
Correlation and Causation
Correlation versus Causation
If the dependent variable increases or decreases when the independent variable increases, there is a positive or negative...
Crossing Over
The homologous pairs of sister chromosomes—one from the maternal and one from the paternal genome—then begin to align alongside each other lengthwise, matching corresponding DNA positions in a process...
Passive Filters
Low-Pass Filters
Low-pass filters are designed to transmit signals with frequencies lower than the cutoff frequency, ωc, and attenuate those above it. The cutoff...
Monohybrid Crosses
Cross-Sectional Research