Permanent Pacemaker Implantation in Patients With Isolated Persistent Left Superior Vena Cava From a Right-Sided
Santosh Kumar Sinha1, Amit Goel1, Mahmodula Razi1
1Department of Cardiology, LPS Institute of Cardiology, G.S.V.M. Medical College, Kanpur, Uttar Pradesh, India.
Background:
Positioning a permanent pacing wire in patients with persistent left superior vena cava (PLSVC) to right ventricle often comes as on-table surprise. It is technically demanding and therefore most of operators prefer left-sided approach. We assessed technical challenges during pacemaker implantation, and their short- and long-term outcomes among patients with isolated PLSVC from a right-sided approach.
Methods:
Thirty-one consecutive patients with isolated PLSVC and 93 patients with right superior vena cava (RSVC) were enrolled with syncope with sinus node dysfunction (SND) and atrioventricular (AV) block. Study was designed on the basis of nested case-control method, and therefore 1:3 proportions was the enrolment criteria to detect any difference as statistically significant as incidence of isolated PLSVC is low.
Results:
Mean age of patients was 64.8 ± 10.5 years. SND was the most common indication (n = 55; 44%) followed by AV block (n = 47; 37%). Nineteen (20%) patients received tined pacing lead, while 105 (85%) had screwing lead. There was no significant difference in mean procedural time (25 ± 11 min vs. 23 ± 12 min; P = 0.24), mean fluoroscopic time (3.1 ± 2.2 min vs. 2.7 ± 2.1 min; P = 0.54), pacing parameters for atrial and ventricular leads, dislodgement rate (3.2% vs. 4.8%; P = 0.32) and follow-up duration (6.9 ± 1.3 years vs. 7.2 ± 1.1 years; P = 0.18) between two groups. Compared to patients with RSVC, those with PLSVC had alpha loop configuration for ventricular lead which was statistically significant (31 vs. 00; P = 0.002).
Conclusions:
Patients with PLSVC had alpha loop configuration for ventricular lead because of circuitous course via left mediastinum. Although pacemaker implantation through coronary sinus via isolated PLSVC from right sided-approach is technically challenging, it obtains good long-term results but needs frequent follow-up during the initial period.
Related Concept Videos
Design Consideration
The factor of safety is another key...
Predicting Reaction Outcomes
Outcomes of Glycolysis
Cellular respiration can occur aerobically (with oxygen) or anaerobically (without oxygen). In the presence of oxygen, cellular respiration starts with glycolysis and continues with pyruvate...
Tensile Strength Considerations of Concrete
The dimensions and shape of a concrete specimen...
Transmission Line Design Considerations
Guidelines for Writing Outcome
Patient outcomes reflect the patient's response to the goal rather than what the nurse aims to achieve. Terminology should be observable and measurable to avoid the reader's interpretation. The desired outcome should be realistic and achievable in the designated care timeframe. Expected outcomes should align with adjunctive therapies. The outcome should enhance care...