Large field of view distortion assessment in a low-field MR-linac
Siamak P Nejad-Davarani1, Joshua P Kim1, Dongsu Du1
1Department of Radiation Oncology, Henry Ford Cancer Institute, 2799 West Grand Blvd., Detroit, MI, 48202, USA.
Purpose:
MR-guided radiation therapy (RT) offers unparalleled soft tissue contrast for localization and target tracking. However, MRI distortions may be detrimental to high precision RT. This work characterizes the gradient nonlinearity (GNL) and total distortions over the first year of clinical operation of a 0.35T MR-linac.
Methods:
For GNL characterization, an in-house large field of view (FOV) phantom (60 × 42.5 × 55 cm3 , >6000 spherical landmarks) was configured and scanned at four timepoints with forward/reverse read polarities (Gradient Echo sequence, FA/TR/TE = 28°/30 ms/6 ms). GNL was measured in Anterior-Posterior (AP), Left-Right (LR), and Superior-Inferior (SI) frequency-encoding directions based on deviation of the auto-segmented landmark centroids between rigidly registered MR and CT images and assessed based on radial distance from magnet isocenter. Total distortion was assessed using a 30 × 30 cm2 grid phantom oriented along the cardinal axes over >1 year of operation.
Results:
The scanner's spatial integrity within the first ~10 months was stable (maximum total distortion variation = 10/6/8%, maximum distortion = 1.41/0.99/1.56 mm in Axial/Coronal/Sagittal planes, respectively). GNL distortions measured during this time period <10 cm from isocenter were (-0.74, 0.45), (-0.67, 0.53), and (-0.86, 0.70) mm in AP/LR/SI directions. In the 10-20 cm range, <1.5% of the distortions exceeded 2 mm in the AP and LR axes while <4% of the distortions exceeded 2 mm for SI. After major repairs and magnet re-shim, detectable changes were observed in total and GNL distortions (20% reduction in AP and 36% increase in SI direction in the 20-25 cm range). Across all timepoints and axes, 38-53% of landmarks in the 20-25 cm range were displaced by >1 mm.
Conclusions:
GNL distortions were negligible within a 10 cm radius from isocenter. However, in the periphery, non-negligible distortions of up to ~7 mm were observed, which may necessitate GNL corrections for MR-IGRT for treatment sites distant from magnet isocenter.
More Related Videos
Related Concept Videos
Electric Field
In the new picture, imagine that the first charge sets up an electric field independent of all other charges in the universe. When another charge comes in its vicinity, the second charge experiences an electric force depending on the electric field at that point. The source charge does not...
Magnetic Fields
A magnetic field is defined by the force that a charged particle experiences...
Electromagnetic Fields
However, the observation of...
Field Effect Transistor
Electric Field Lines
The solution to this problem is to use electric field lines, which are not vectors but...
Magnetic Field Lines
Magnetic field lines follow several hard-and-fast rules: