Validation of finger blood pressure monitoring in children
Natalie D Heeney1, Fajish Habib2, Garveen K Brar1
1Departments of Biomedical Physiology and Kinesiology.
Insights
Noninvasive finger blood pressure monitoring using reconstructed finger-brachial arterial pressure (reBAP) is a valid method for children. This approach offers a comfortable, convenient, and accurate alternative for continuous hemodynamic monitoring in pediatric patients.
Area of Science:
- Pediatric Cardiology
- Biomedical Engineering
- Hemodynamics
Background:
- Continuous beat-to-beat blood pressure monitoring is crucial for cardiovascular reflex testing and hemodynamic monitoring.
- Noninvasive finger blood pressure monitoring is accurate in adults but lacks validation in children.
- Invasive intra-arterial blood pressure monitoring is the gold standard but is invasive.
Purpose of the Study:
- To evaluate the accuracy of novel noninvasive finger arterial pressure (FinAP) measurements in children.
- To compare FinAP measurements with invasive intra-arterial blood pressure monitoring.
Main Methods:
- Used prototype pediatric finger cuffs for noninvasive measurements.
- Compared raw FinAP and reconstructed finger-brachial arterial pressure (reBAP) against invasive intra-arterial blood pressure.
- Utilized Bland-Altman analyses, cumulative percentage differences (clinical grades A-D), and waveform morphology analysis.
Main Results:
- Analyzed data from 13 children (aged 3-13 years).
- reBAP showed bias of 1.8±6.9 mmHg (systolic), 0.3±6.1 mmHg (diastolic), and 0.4±5.3 mmHg (mean arterial pressure).
- Clinical grades for reBAP were C, B, and A for systolic, diastolic, and mean arterial pressure, respectively; reBAP improved numerical accuracy but reduced waveform agreement.
Conclusions:
- Middle finger arterial pressure measurements with waveform reconstruction are an acceptable surrogate for invasive monitoring in children.
- Finger blood pressure monitoring offers a novel, comfortable, convenient, and accurate noninvasive alternative for pediatric beat-to-beat blood pressure monitoring.
Background:
Continuous beat-to-beat blood pressure monitoring permits the rapid detection of blood pressure fluctuations for cardiovascular reflex testing and clinical haemodynamic monitoring. In adults, this can be achieved noninvasively with high accuracy, using finger blood pressure monitoring with volume clamp photoplethysmography. However, data are lacking on the validity of finger blood pressure monitoring in children compared to the gold standard - invasive intra-arterial blood pressure monitoring.
Aim:
We aimed to evaluate the accuracy of novel noninvasive index and middle finger arterial pressure (FinAP) measurements in children.
Methods:
Using prototype paediatric finger cuffs, we compared: mean differences, bias and limits of agreement (Bland-Altman analyses); cumulative percentage differences [clinical grade A-D (based on the percentage of heartbeats in agreement with the standard)]; and waveform morphology (regression analysis and smoothing) between both raw FinAP (Finapres NOVA) and reconstructed finger-brachial arterial pressure (reBAP) compared to intra-arterial blood pressure measurements.
Results:
Eighteen children were tested (aged 3-13 years; 12 male), with data from 13 included in the analysis. The bias for reBAP for the middle finger was 1.8±6.9, 0.3±6.1 and 0.4±5.3 mmHg for systolic, diastolic and mean arterial pressure, with clinical grades of C, B and A, respectively. reBAP improved numerical accuracy, but reduced waveform morphological agreement.
Conclusion:
Middle finger arterial measurements with waveform reconstruction provide an acceptable surrogate for invasive intra-arterial recording in children. Finger blood pressure monitoring is a novel comfortable, convenient and accurate alternative approach for noninvasive beat-to-beat blood pressure monitoring in children.
More Related Videos
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Blood Pressure
Blood Pressure
The average BP in an adult is typically around 120/80 mmHg (millimeters of mercury). In this measurement, the numerator (120) indicates the systolic pressure, which is the pressure in the arteries during the contraction of the heart's ventricles as blood is expelled. The denominator (80) represents the...
Measurement of Blood Pressure
Factors affecting Blood pressure
Physiological Factors:
Alterations in Blood Pressure
Hypertension (High blood pressure)
Hypertension occurs when blood pressure readings consistently exceed the normal range. It is diagnosed when systolic blood pressure (the top number, indicating pressure while the heart...