Evaluation of sacral neuromodulation system with new six-contact points electrode in pigs
Yiming Wang1,2,3, Limin Liao1,2,3, Huiling Cong1,2,3
1Department of Urology of Capital Medical University, China Rehabilitation Research Center, Capital Medical University, Beijing, China.
Aims:
We investigated the effects of sacral neuromodulation using the new six-contact electrode vs the four-contact electrode in pigs.
Methods:
Randomly, a six-contact electrode was implanted in eight pigs in one side of the third sacral (S3) foramen, and a four-contact electrode was implanted in the other side using the same method. Using an external neurostimulator, the number of contact points (sensitive voltage ≤ 2 V) of both electrodes (SacralStim and InterStim systems) was calculated. Cystometry was performed by infusing normal saline or acetic acid. Then sacral neuromodulation with the SacralStim and InterStim systems was induced at a voltage at which we could observe perianal and/or tail movement. Multiple cystometrograms were performed to determine the effects of the two systems on the micturition reflex.
Results:
The mean number of sensitive points of six-contact electrodes of the SacralStim system (2.63 ± 0.32) was higher than that of the quadripolar-lead electrodes of the InterStim system (1.38 ± 0.18), and the difference was statistically significant (P < 0.05). Acetic acid-induced bladder overactivity significantly reduced bladder capacity to 54.89% ± 4.7% of the normal saline control level. During acetic acid infusion, sacral neuromodulation with the SacralStim system suppressed bladder overactivity and significantly increased bladder capacity to 70.41% ± 5.4% of the normal saline control level, compared with the acetic acid level ( P < 0.05). Moreover, sacral neuromodulation with the InterStim system also significantly increased bladder capacity to 69.63% ± 5.3% of the normal saline control level, compared with the acetic acid level ( P < 0.05). No significant differences were found in the results obtained using the two systems ( P > 0.05).
Conclusions:
The six-contact electrode of the SacralStim system had more sensitive points (<2 V) than that of the quadripolar-lead electrode of InterStim system. Potentially, it has more postimplantation programming options and battery savings manifested by lower voltage will increase the longevity of the stimulator. Further studies of sacral neuromodulation with six-contact electrodes in clinical practice are needed.
Related Concept Videos
Contact-dependent Signaling
Gap Junctions
In animal cells, gap junctions are formed...
Contact Angle
The adhesive force is the molecular force between molecules of different materials, that is, between the molecules of the solid and the liquid. The cohesive...
Standard Electrode Potentials
Electrodes: Overview
There are two main types of electrodes in electrochemical cells. The first type, known as the working or indicator electrode, has a potential that is sensitive to the analyte's concentration and reacts to changes in...
Potentiometry: Types of Electrodes
The Standard Hydrogen Electrode (SHE) is a widely used reference electrode that maintains zero potential across all temperatures. However, its need for a continuous hydrogen gas supply renders it impractical for everyday use.
An alternative to SHE is the Saturated Calomel Electrode (SCE). This electrode features an...
Potentiometry: Membrane Electrodes