Murine Model of Intestinal Ischemia-reperfusion Injury
Published on: May 11, 2016
Intestinal Injury by Heat Conduction from Surgical Sealing Devices
Toshiro Suzuki1, Ryouhei Hattori2, Tomonori Minagawa1
1Department of Urology, Shinshu University School of Medicine, Matsumoto, Japan.
Background:
There have been no investigations of intestinal injury induced by surgical sealing devices, especially focusing heat conduction from the back of active blades during laparoscopic surgery.
Objective:
This study of damage to the small intestine by heat conduction from the back of active blades both physically and histopathologically was performed to establish safe usage of surgical sealing devices.
Materials And Method:
We compared seven types of bipolar sealing device and two types of ultrasonic coagulating shear in an animal model simulating laparoscopic surgery. Time-dependent changes in heat conduction from the back of active blades were measured using a direct contact thermometer during intracorporeal activation. Histopathological damage to the small intestine by the back of active blades in laparoscopic surgical application was evaluated. The backs of active blades were activated while attached to the serosa of the small intestine. The depths of histopathological changes were measured to evaluate the thermal effects of surgical sealing devices.
Results:
Most devices generated temperatures >70°C even on the back of active blades. There were no significant differences in duration for cooling to ≤50°C among these devices. All devices induced histopathological heat damage in the submucosal layer or deeper.
Conclusions:
Regardless of type, the backs of active blades of surgical sealing devices conduct high temperatures and can induce heat damage in the small intestine. Surgical sealing devices should not be activated while attached to surrounding tissue or organs in laparoscopic surgery.
Related Concept Videos
Anatomy of the Intestines
Small Intestines
The small intestine is an ~7 meter-long tube with an inner diameter of just 2.5 cm. Since most nutrients are absorbed here, the inner lining of the...
Quantifying Heat
Specific Heat
For example, increasing the temperature of one gram of water by 1°C requires one calorie of heat energy and can be written as 1 cal/g-°C, or...
Heat Flow and Specific Heat
Conduct Disorder
Conduction System of the Heart
The pacemaker cells are located in two primary nodes: the sinoatrial (SA) node and the atrioventricular (AV) node. The SA node pacemaker cells can autonomously depolarize, triggering an action potential that leads to the...