Related Experiment Video
Updated: Jan 28, 2026

Investigating the Potential of Singly Curved Thin Piezoelectric Transducers for Energy Harvesting and Structural Health Monitoring
Published on: November 14, 2025
Best zero level for external ICP transducer
Peter Reinstrup1, Mårten Unnerbäck2, Niklas Marklund3
1Department of Intensive & Perioperative care, Skanes University Hospital, Lund, Sweden. Peter.Reinstrup@med.lu.se.
Background:
Continuous monitoring of intracranial pressure (ICP) was introduced in the 1950s. For correct ICP recordings, the zero-reference point for the external pressure gauge must be placed next to a head anatomical structure. We evaluated different anatomical points as zero reference for the ICP device at different head positions and their relation to brain centre (BC), foramen of Monro (Monro), and brain surface.
Methods:
Patients referred for neuroimaging due to e.g. headache all having normal 3D MRI scans were selected. Monro, BC, Orbit(O), external auditory meatus (EAM), and orbito-meatal (OM) line were identified and projected to mid-sagittal, or axial images. Each scan was evaluated like lying supine, 45° head elevations, upright, and 45° lateral position. Distances from skin to brain surface, BC, and Monro were measured. All values are presented as mean ± SD and/or range in millimetre. For conversion to mmHg, millimetre was multiplied by 0.074.
Results:
Twenty MRI scans were examined. A zero reference at EAM or glabella was ideal at BC when head was strict supine or in the lateral position. At 45° head elevation, an overestimation of the BC-ICP by 4.8 ± 0.8 and in upright 5.6 ± 0.5 mmHg was found, and 45° lateral underestimated ICP-BC by 6.3 ± 1.0 mmHg. Monro was situated 45 ± 5 mm rostral to the mid-OM line and 24 (18-31) mm inferior and 13 (8-17) mm in front of BC. A zero-reference point aligned with the highest point of the head underestimated BC-ICP and Monro-ICP. If the ICP reading was added 5.9 or 6.3 mmHg, respectively, a deviation from BC-ICP was ≤ 1.8 mmHg and Monro-ICP was ≤ 0.9 mmHg in all head positions.
Conclusions:
EAM and glabella are defined anatomical structures representing BC when strict supine or lateral but with 12 mmHg variation with different head positions used in clinical practice. The OM line follows Monro at head elevation, but not when the head is turned. When the highest external point on the head is used, ICP values at brain surface as well as Monro and BC are underestimated. This underestimation is fairly constant and, when corrected for, provides the most exact ICP reading.
Related Concept Videos
Work Done on a System by External Force
In the presence of a non-conservative opposing force, like friction, some part of the work done...
External and Internal Respiration
Internal and External Forces
External Anatomy of the Kidney
The kidneys are located in the retroperitoneal space on either side of the vertebral column, protected posteriorly by the 11th and 12th ribs. The right kidney sits slightly lower than the left owing to the presence of the liver...
External Female Genitals
General External Flow Characteristics

