Fretting and Corrosion Damage of Retrieved Dual-Mobility Total Hip Arthroplasty Systems
Daniel J Lombardo1, Matthew P Siljander1, Corinn K Gehrke2
1Department of Orthopaedic Surgery, Beaumont Health, Royal Oak, MI.
Background:
Dual-mobility (DM) total hip arthroplasty (THA) systems are designed to increase stability while potentially avoiding problems associated with large femoral heads. Complications of these systems are not yet fully understood. This study aims at characterizing in vivo performance of DM hip systems and assessing modes of clinical failure.
Methods:
Under an institutional review board-approved implant retrieval protocol, 18 DM THA systems from 17 patients were included. Implants were graded at the head-neck junction for fretting and corrosion based on the system of Goldberg et al. Components were also macroscopically examined for different damage modes. Demographics and surgical data were collected from medical records, and radiographs were assessed for component positioning. Data were analyzed through Spearman rank-order correlation and Mann-Whitney U-tests, with α = 0.05.
Results:
The average length of implantation was 13.4 months with mild to moderate fretting corrosion damage. Polyethylene (PE) liners exhibited edge deformation, scratching, and pitting damage. Metallic components exhibited burnishing and scratching damage. Summed fretting and corrosion scores were strongly correlated (ρ = 0.967, P < .0001). Summed corrosion score was moderately correlated with presence of embedding on the PE liner (ρ = 0.690, P = .017). PE liner abrasion and edge deformation of the femoral stem taper were moderately positively correlated (ρ = 0.690, P = .017). Fretting and corrosion damage were not significantly correlated with patient demographics or radiographic positioning of implants. There were no differences in scores between modular and monoblock designs.
Conclusion:
These findings demonstrate that dual-mobility THA systems may be susceptible to the same fretting and corrosion damage observed in traditional modular THA systems. Future studies are needed to confirm these results and clinical significance.
Related Concept Videos
Corrosion
Retrieval
Recall involves accessing information without cues, such as during an essay test, where individuals must retrieve facts and concepts from memory unaided. Another example is remembering the name of a colleague...
Corrosion of Reinforcement
However, over time and under certain conditions like carbonation, chloride ingress, and cracking this protective state can be compromised. Steel has areas with...
ER Retrieval Pathway
The ER uses many checkpoints to prevent the entry of incorrectly folded or a resident protein as cargo onto a transport vesicle. These mechanisms...
Second Order systems II
First Order Systems
When a first-order system is subjected to a unit-step input, its response is characterized by its transfer function. By applying the Laplace transform of the unit-step input to the transfer function, expanding the...