Related Experiment Video
Updated: Jan 28, 2026

Myocardial Infarction and Functional Outcome Assessment in Pigs
Published on: April 25, 2014
Comparative outcomes of subcutaneous and transvenous cardioverter-defibrillators
Jin-Jun Liang1,2, Hideo Okamura2, Roshini Asirvatham2
1Department of Cardiology, Renmin Hospital of Wuhan University, Wuhan, Hubei 430060, China.
Background:
The comparative outcomes of subcutaneous implantable cardioverter-defibrillator (S-ICD) and transvenous ICD (T-ICD) have not been well studied. The aim of this study was to evaluate the safety and efficacy of currently available S-ICD and T-ICD.
Methods:
The study included 86 patients who received an S-ICD and 1:1 matched to those who received single-chamber T-ICD by gender, age, diagnosis, left ventricular ejection fraction (LVEF), and implant year. The clinical outcomes and implant complications were compared between the two groups.
Results:
The mean age of the 172 patients was 45 years, and 129 (75%) were male. The most common cardiac condition was hypertrophic cardiomyopathy (HCM, 37.8%). The mean LVEF was 50%. At a mean follow-up of 23 months, the appropriate and inappropriate ICD therapy rate were 1.2% vs. 4.7% (χ = 1.854, P = 0.368) and 9.3% vs. 3.5% (χ = 2.428, P = 0.211) in S-ICD and T-ICD groups respectively. There were no significant differences in device-related major and minor complications between the two groups (7.0% vs. 3.5%, χ = 1.055, P = 0.496). The S-ICD group had higher T-wave oversensing than T-ICD group (9.3% vs. 0%, χ = 8.390, P = 0.007). Sixty-five patients had HCM (32 in S-ICD and 33 in T-ICD). The incidence of major complications was not significantly different between the two groups.
Conclusions:
The efficacy of an S-ICD is comparable to that of T-ICD, especially in a dominantly HCM patient population. The S-ICD is associated with fewer major complications demanding reoperation.
Related Concept Videos
Predicting Reaction Outcomes
Outcomes of Glycolysis
Cellular respiration can occur aerobically (with oxygen) or anaerobically (without oxygen). In the presence of oxygen, cellular respiration starts with glycolysis and continues with pyruvate...
Guidelines for Writing Outcome
Patient outcomes reflect the patient's response to the goal rather than what the nurse aims to achieve. Terminology should be observable and measurable to avoid the reader's interpretation. The desired outcome should be realistic and achievable in the designated care timeframe. Expected outcomes should align with adjunctive therapies. The outcome should enhance care...
Comparing Copy Number Variations and SNPs
Copy number variations or CNVs are the structural variations that cover more than 1kb of DNA sequence. The single nucleotide polymorphism (SNP), on the other hand, is a single nucleotide change or a point mutation that is found in more than 1%...
Comparing the Survival Analysis of Two or More Groups
Comparing Experimental Results: Student's t-Test