Intraoperative High Impedance Levels During Placement of Deep Brain Stimulating Electrode
Mark K Lyons1, Matthew T Neal1, Naresh P Patel1
1Department of Neurological Surgery Mayo Clinic Arizona Phoenix, Arizona.
Background And Importance:
Deep brain stimulation (DBS) is a well-established treatment for medically refractory Parkinson's disease (PD), essential tremor (ET), and dystonia. The field of DBS is expanding and techniques are under investigation for the treatment of several neurological disorders. A critical component of the success of these procedures depends significantly on the reliability and durability of devices implanted. Immediate feedback during surgery often gives the surgeon and patient a sense of confidence of long term success. When impedances are found to be elevated during the implantation of the DBS leads, appropriate trouble shooting measures are critical.
Clinical Presentation:
We present a 73-yr-old male undergoing awake subthalamic DBS with microelectrode recordings for severe PD. Once the optimal trajectory and depth were ascertained, the permanent DBS electrode was placed. High impedances were recorded. Troubleshooting procedures were performed and were all negative as to the cause of the values. Correct impedance levels of the DBS electrode was confirmed with extracranial testing, but continued high values were found again with intracranial positioning of the electrode. A postoperative computerized tomography (CT) scan confirmed intracranial air surrounding all of the contacts. The patient went on to outpatient programming with excellent clinical results.
Conclusion:
The presence of pneumocephalus surrounding the DBS lead contacts at the target nucleus may have accounted for the intraoperative impedance findings. When all troubleshooting checks have not identified an explanation for the high impedances, intraoperative imaging may demonstrate pneumocephalus around the lead contacts, which should resolve and impedances return to normal values.
More Related Videos
09:16A General Method for Evaluating Deep Brain Stimulation Effects on Intravenous Methamphetamine Self-Administration
Published on: January 22, 2016
10:52Microelectrode Guided Implantation of Electrodes into the Subthalamic Nucleus of Rats for Long-term Deep Brain Stimulation
Published on: October 2, 2015
Related Concept Videos
Standard Electrode Potentials
Impedance Combination
Impedances and Admittance
Impedance is a measure of resistance to sinusoidal current flow in an AC circuit. Unlike their behavior in DC circuits, where inductors appear as short circuits and capacitors as open circuits, the behavior of these components in AC...
Series Impedances: Three-Phase Line
Using Kirchhoff's laws, an integro-differential equation for the network is derived. This equation accounts for unbalanced phase currents, which may induce return currents through neutral wires and the earth, seeking the least impedance path. Earth return conductors can replace the...
Bus Impedance Matrix
In the first circuit, all machine voltage sources are short-circuited, leaving only the prefault voltage source at the fault location. The positive-sequence bus impedance matrix can be determined by solving the nodal equations,...
Line Protection with Impedance Relays
Under normal conditions, low load currents keep the measured...