Background:
Interruptions in compressions reducethe efficacy of cardiopulmonary resuscitation (CPR) and are inevitable during hands-off periods for shocks. Clinical exam gloves were found to facilitate safe contact with patients during shock delivery but the safety of this practice has been questioned. Polyethylene is of interest because of its safety record in the medical arena and its electrical insulation properties.
Methods:
This study measured the current leak through 2 mil (0.002 inch) polyethylene drapes during shock delivery. The current leak was assessed by measurement of voltage changes in a circuit recommended by the International Electrotechnical Commission (IEC) for current leak safety testing. Current flowed off the drape, through the circuit and to electric ground in a manner consistent with standardized testing. Perceptibility was assessed in a subset with the investigator's bare hands pressed into the drape during shock delivery.
Results:
Thirty-three patients undergoing elective cardioversion at Emory University Hospital underwent analysis (age 23-90, 36% female). Biphasic energies were 200-360 J. The root mean square (RMS) current leak averaged 0.072 ± 0.022 mA and peak current leak averaged 0.67 ± 0.21 which is well below IEC recommendations of 3.5 mA RMS and 5.0 mA peak. Finally, no instances of dielectric breakdown occurred and no shocks were perceptible.
Conclusions:
Polyethylene is a common medical material which may facilitate safe hands-on defibrillation. Our data illustrates that a thin, semitransparent layer of polyethylene is a safe and feasible adjunct to cardiac arrest kits to allow continued compressions and simplification of the CPR process.
More Related Videos
05:52Handwriting Analysis Indicates Spontaneous Dyskinesias in Neuroleptic Naïve Adolescents at High Risk for Psychosis
Published on: November 21, 2013
09:35Measurement of the Hand Transmitted Vibration of the Human Hand Arm System During Operation of a Hand Tractor
Published on: June 16, 2021
Related Concept Videos
Survey Safety
Hand hygiene
Hand washing...
Household Wiring And Electrical Safety
Relative Risk
Physiological Barriers
The blood endothelial barrier is the most porous of these. It allows all small ionized, un-ionized, and lipophilic molecules to pass through the endothelial lining into the interstitial space...
Potential Energy
Chemical bonds that form attractive forces between atoms also contain potential energy, called chemical energy. When a chemical reaction...