Robust 3D surface recovery by applying a focus criterion in white light scanning interference microscopy

Applied Optics
|March 16, 2019
PubMed
Summary

We introduce Shape from Focus-White Light Scanning Interference (SFF-WLSI) microscopy to improve 3D surface topography measurements. This new technique accurately measures rough or low-reflectivity surfaces, outperforming traditional methods.

Related Concept Videos

Focusing of Light in the Eye01:16

Focusing of Light in the Eye

Light rays enter the eye through the cornea, a transparent dome-shaped tissue that is the eye's outermost layer. The cornea bends or refracts, light rays traveling to the pupil. The shape of the cornea determines how much of the light is bent and whether the image will be focused correctly on the retina at the back of the eye. Once the light has passed through both refraction layers, it converges into a single focal point onto a small area. This is where photoreceptors start transforming...
5.8K
Scanning Electron Microscopy01:07

Scanning Electron Microscopy

A scanning electron microscope (SEM) is used to study the surface features of a sample by using an electron beam that scans the sample surface in a two-dimensional manner. Typically, areas between ~1 centimeter to 5 micrometers in width can be imaged. SEM can be used to image bacteria, viruses, tissues as well as larger samples like insects. Conventional SEM gives a magnification ranging from 20X to 30,000X and spatial resolution of 50 to 100 nanometers.
Fundamental Principles
Accelerated...
5.5K
Interference and Diffraction02:18

Interference and Diffraction

Interference is a characteristic phenomenon exhibited by waves. When two electromagnetic waves interact with their peaks and troughs coinciding, a resulting wave with enhanced amplitude is produced. This is known as constructive interference. In this case, the two waves interacting are in phase with each other.
52.3K
Phase Contrast and Differential Interference Contrast Microscopy01:26

Phase Contrast and Differential Interference Contrast Microscopy

Phase-Contrast Microscopes
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
13.2K
Routh-Hurwitz Criterion I01:15

Routh-Hurwitz Criterion I

Consider an electrical power grid, where stability is essential to prevent blackouts. The Routh-Hurwitz criterion is a valuable tool for assessing system stability under varying load conditions or faults. By analyzing the closed-loop transfer function, the Routh-Hurwitz criterion helps determine whether the system remains stable.
To apply the Routh-Hurwitz criterion, a Routh table is constructed. The table's rows are labeled with powers of the complex frequency variable s, starting from the...
596
Routh-Hurwitz Criterion II01:19

Routh-Hurwitz Criterion II

In the application of the Routh-Hurwitz criterion, two specific scenarios can arise that complicate stability analysis.
The first scenario occurs when a singular zero appears in the first column of the Routh table. This situation creates a division by zero issues. To resolve this, a small positive or negative number, denoted as epsilon (∈), is substituted for the zero. The stability analysis proceeds by assuming a sign for ∈. If ∈ is positive, any sign change in the first...
1.0K