Biomechanical comparison of 2 augmented glenoid designs: an integrated kinematic finite element analysis
Vani J Sabesan1, Diego J L Lima1, James D Whaley2
1Department of Orthopedic Surgery, Cleveland Clinic Florida, Weston, FL, USA.
Background:
Augmented glenoid implants are available to help restore the biomechanics of the glenohumeral joint with excessive retroversion. It is imperative to understand their behavior to make a knowledgeable preoperative decision. Therefore, our goal was to identify an optimal augmented glenoid design based on finite element analysis (FEA) under maximum physiological loading.
Methods:
FEA models of 2 augmented glenoid designs-wedge and step-were created per the manufacturers' specifications and virtually implanted in a scapula model to correct 20° of retroversion. Simulation of shoulder abduction was performed using the FEA shoulder model. The glenohumeral force ratio, relative micromotion, and stress levels on the cement mantle, glenoid vault, and backside of the implants were compared between the 2 designs.
Results:
The force ratio was 0.56 for the wedge design and 0.87 for the step design. Micromotion (combination of distraction, translation, and compression) was greater for the step design than the wedge design. Distraction measured 0.05 mm for the wedge design and 0.14 mm for the step component. Both implants showed a similar pattern for translation; however, compression was almost 3 times greater for the step component. Both implants showed high stress levels on the cement mantle. At the glenoid vault and on the implants, the stress levels were 1.65 MPa and 6.62 MPa, respectively, for the wedge design and 3.78 MPa and 13.25 MPa, respectively, for the step design.
Conclusion:
Implant design slightly affects joint stability; however, it plays a major role regarding long-term survival. Overall, the augmented wedge design provides better implant fixation and stress profiles with less micromotion.
More Related Videos
Related Concept Videos
Periodic Classification of the Elements
Classification of Elements and Compounds
Compounds are pure substances composed of two or more elements in fixed, definite proportions. Compounds are classified as ionic or molecular (covalent) based on the bonds...
Elements and Compounds
Elements
Elements are classified as atomic or molecular based on the nature of their basic units. They are unique forms of matter with specific chemical and physical properties that cannot break down into smaller substances by ordinary chemical reactions. There...
Elements: Chemical Symbols and Isotopes
Some symbols are derived from the common English name of the element; others are abbreviations of the name in another language — Latin, Greek or German. For example, the symbol for aluminum (common name)...
Kinematic Equations - I
Key Elements for Plant Nutrition