Related Experiment Video
Updated: Jan 27, 2026

X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
X-ray examination dose surveys: how accurate are my results?
Stephen Taylor1, Alain Van Muylem2, Nigel Howarth3
1Department of Radiology, Hôpital Ambroise Paré, Boulevard Président Kennedy 2, 7000, Mons, Belgium.
Objectives:
To determine the variabilities of dose-area-products (DAP) of frequent X-ray examinations collected for comparison with diagnostic reference levels (DRLs).
Methods:
DAP values of chest, abdomen, and lumbar spine examinations obtained on devices from two manufacturers were collected in three centers over 1 to 2 years. The variability of the average DAP results defined as the 95% confidence interval in percentage of their median value was calculated for increasing sample sizes, each examination and center. We computed the sample sizes yielding variabilities lower or equal to 25% and 10%. The effect of narrowing patient selection based on body weight was also investigated (ranges of 67-73 Kg, or 60-80 Kg).
Results:
DAP variabilities ranged from 75 to 170% of the median value when collecting small samples (10 to 20 DAP). To reduce this variability, larger samples are needed, collected over up to 2 years, regardless of the examination and center. A variability ≤ 10% could only be reached for chest X-rays, requiring up to 800 data. For the abdomen and lumbar spine, the lowest achievable variability was 25%, regardless of the body weight selection, requiring up to 400 data.
Conclusion:
Variabilities in DAP collected through small samples of ten data as recommended by authorities are very high, but can be reduced down to 25% (abdomen and lumbar spine) or even 10% (chest) through a substantial increase in sample sizes. Our findings could assist radiologists and regulatory authorities in estimating the reliability of the data obtained when performing X-ray dose surveys.
Key Points:
• Low but reasonable variabilities cannot be reached with samples sized as recommended by regulatory authorities. Higher numbers of DAP values are required to reduce the variability. • Variabilities of 10% for the chest and 25% for abdomen and lumbar spine examinations are achievable, provided large samples of data are collected over 1 year. • Our results could help radiologists and authorities interpret X-rays dose surveys.
Related Concept Videos
Surveys
Introduction to Surveying, Plane Surveying and Geodetic Surveys
X-ray Crystallography
Diffraction
Diffraction is the change in the direction of travel experienced by an electromagnetic wave when it encounters a physical barrier whose dimensions are comparable to those of the wavelength of the light. X-rays are electromagnetic radiation with wavelengths about as long as the distance between neighboring...
Types of Surveys
Survey Safety
Errors and Mistakes in Surveying