Determination of homogeneity of the top surface deadlayer in an old HPGe detector
Mikael Hult1, Stef Geelen2, Mark Stals2
1European Commission, Joint Research Centre, JRC-Geel, Retieseweg 111, B-2440, Geel, Belgium.
Abstract:
A collimated source of 241Am was scanned over the endcap of a 21 year old coaxial HPGe-detector that had spent about 75% of its life at room temperature (and the remaining time at 77 K). The detector response was recorded and used as a measure of the relative thickness of the top deadlayer. This thickness was not homogeneous and was thicker near to the outer surface of the crystal compared to the centre, which could be a result of increased diffusion of Li atoms during times the detector was kept at room temperature. The results were compared with two newer HPGe-detectors that proved to have homogeneous top deadlayers.
More Related Videos
08:54Determining Four Components in a Lipid Nanoparticle RNA Delivery System by Liquid Chromatography Combined with Evaporative Light Scattering Detector
Published on: May 30, 2025
06:53Author Spotlight: Technologies and Challenges in Elemental Analysis of Food Samples
Published on: December 22, 2023
Related Concept Videos
Test for Homogeneity
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography: Types of Detectors-II
Homogeneous Equilibria for Gaseous Reactions
For gas-phase reactions, the equilibrium constant may be expressed in terms of either the molar concentrations (Kc) or partial pressures (Kp) of the reactants and products. A relation between these two K values may be simply derived from the ideal gas equation and the definition of molarity. According to the ideal gas equation:
High-Performance Liquid Chromatography: Types of Detectors