Commissioning of portal dosimetry using a novel method for flattening filter-free photon beam in a nontrue beam
Arun Gandhi1, Subramani Vellaiyan2, V S Subramanian1
1Department of Radiation Oncology, Yashoda Hospitals, Hyderabad, Telangana; Research and Development Centre, Bharathiar University, Coimbatore, Tamil Nadu, India.
Aim:
The aim of this study is to commission and validate the portal dosimetry (PD) system using an indirect method for flattening filter free (FFF) photon beam of the upgraded c-series linear accelerator.
Background:
Varian Medical System clinacs with amorphous-silicon portal imager panel (aSi-1000) do not have PD for FFF beams. Recently, our c-series linear accelerator was upgraded to deliver 6MV FFF (6MVFFF) photon beam with the highest dose rate of 1400 monitor unit (MU)/min. The study, therefore, focuses on the commissioning and validation of PD for the 6MVFFF beam.
Materials And Methods:
An indirect method was implemented to predict the portal dose for FFF beam in Eclipse as the treatment planning system does not have direct prediction algorithm for FFF beam (version. 11). Dosimetrical characteristics of aSi-electronic portal imaging device (EPID) were evaluated for 6MVFFF beam and validation of PD for 6MVFFF beam was performed for open fields along with pretreatment quality assurance of intensity-modulated radiation therapy (IMRT), volumetric-modulated arc therapy (VMAT), and stereotactic radiosurgery (SRS) techniques for 30 patients planned with 6MVFFF beam.
Results:
ASi-EPID saturates between 100 and 130 cm source to detector distance (SDD) for 6MVFFF beam and resolved at more than 140 cm SDD. The squared correlation coefficient (R2) for MU linearity was found to be 1 (R2 = 1), and instantaneous dose response linearity at different SDD's was found to be 0.999 (R2 = 0.999) for the 6MVFFF beam. Maximum gamma area index (GAI) for 3% dose difference and 3 mm distance-to-agreement criteria for IMRT, VMAT, and SRS/stereotactic radiotherapy plans was 97.9% ± 0.3%, 96.3% ± 0.5%, and 98.2% ± 0.2%, respectively.
Conclusion:
The results reveal that this novel method can be used to commission portal dosimetry for 6MVFFF beam as it is a convenient, faster, and accurate method.
Related Concept Videos
Beams
Based on geometry, beams can be straight, tapered, or curved. Straight beams are the most common type and have a constant cross-section throughout their length. Tapered beams, on the other hand, have a varying cross-section along...
Deflection of a Beam
Singularity functions, described in an earlier lesson, are powerful mathematical tools that represent discontinuities within a function commonly encountered in structural loading...
Prismatic Beams: Problem Solving
The design begins with analyzing the beam as a free body to identify moments and force balances, thereby determining support reactions. Next, the...
Principal Stresses in a Beam
Analyzing principal stresses is crucial, especially in...
Beams with Symmetric Loadings
The M/EI...
Beams with Unsymmetric Loadings
The first moment-area theorem determines the slope at any point on the beam. This theorem indicates that the change in slope between two points on a beam...