Taper Corrosion and Adverse Local Tissue Reactions in Patients with a Modular Knee Prosthesis
Tom Christiner1,2, Moreica B Pabbruwe1, Alan M Kop1
1Centre for Implant Technology and Retrieval Analysis (CITRA), Department of Medical Engineering and Physics, Royal Perth Hospital, Perth, Australia.
Corrosion in modular knee implants is linked to adverse local tissue reactions and osteolysis. This study highlights fretting corrosion at modular junctions as a cause, urging clinician awareness for earlier detection and management.
Area of Science:
- Orthopedic Surgery
- Biomaterials Science
- Pathology
Background:
- Corrosion is observed in modular knee implants but its link to patient outcomes is unclear.
- This study investigated fretting corrosion at modular junctions in Stryker Triathlon TS knee implants.
- Three of 13 retrieved implants showed osteolysis and adverse tissue reactions due to fretting corrosion.
Purpose of the Study:
- To investigate the prevalence and severity of corrosion in modular knee implant components.
- To identify factors influencing corrosion development.
- To correlate corrosion with adverse local tissue reactions and implant failure.
Main Methods:
- Retrieval analysis of 13 Stryker Triathlon TS modular knee implants.
- Examination of modular surfaces for corrosion using scanning electron microscopy and energy-dispersive x-ray spectroscopy.
- Histopathological analysis of tissue samples from patients with adverse local tissue reactions.
Main Results:
- Corrosion was present in 62% of tibial and 75% of femoral components.
- Tibial corrosion was less prevalent but occurred earlier and was more severe.
- Histopathology revealed severe reactions to metal debris, including aseptic lymphocyte-dominated vasculitis-associated lesion (ALVAL).
Conclusions:
- Aseptic lymphocyte-dominated vasculitis-associated lesion (ALVAL) and pseudotumors are reported secondary to modular knee replacement corrosion.
- The threaded taper design and release of cobalt-chromium ions/debris are implicated in adverse reactions, osteolysis, and tissue damage.
- Clinicians must be aware of this complication for earlier recognition and management.
Related Concept Videos
Corrosion
Local Anesthetics: Adverse Effects
Once absorbed into the systemic circulation, local anesthetics can affect the organs that depend on the functioning of sodium...
Corrosion of Reinforcement
However, over time and under certain conditions like carbonation, chloride ingress, and cracking this protective state can be compromised. Steel has areas with...
Knee Joint
A total of seven ligaments support the knee joint. The patellar ligament, which is also attached to the quadriceps femoris...
Oxidation-Reduction Reactions
Tissue-Drug Binding: Localization of Drugs and its Significance
Drugs can bind to different tissue components, enhancing their distribution and localization. The factors influencing drug localization in tissues include the drug's lipophilicity, structural characteristics, tissue perfusion rate, and pH differences. These factors determine...