Current assessment of pulse wave velocity: comprehensive review of validation studies
Alberto Milan1, Gaia Zocaro1, Dario Leone1
1Internal Medicine and Hypertension Division, Department of Medical Sciences, University of Torino, AOU Città della Salute e della Scienza di Torino.
Objective:
Carotid-femoral pulse wave velocity (PWV) is considered the gold standard for arterial stiffness assessment in clinical practice. A large number of devices to measure PWV have been developed and validated. We reviewed different validation studies of PWV estimation techniques and assessed their conformity to the Artery Society Guidelines and the American Heart Association recommendations.
Methods:
Pubmed and Medline (1995-2017) were searched to identify PWV validation studies. Of the 96 article retrieved, 26 met the inclusion criteria.
Results:
Several devices had been developed and validated to noninvasively measure arterial stiffness, using applanation tonometry (SphygmoCor, PulsePen), piezoelectric mechanotransducers (Complior), cuff-based oscillometry (Arteriograph, Vicorder and Mobil-O-Graph), photodiode sensors (pOpmètre) and devices assessing brachial-ankle pulse wave velocity and cardiac-ankle PWV. Ultrasound technique and MRI remain confined to clinical research. Good agreement was found with the Artery Society Guidelines. Two studies (Complior, SphygmoCor Xcel) showed best adherence with the guidelines. In Arteriograph, MRI, ultrasound and SphygmoCor Xcel validation studies sample size was smaller than the minimum suggested by the guidelines. High discrepancies between devices were shown in distance estimation: in two studies (Arteriograph, Complior) path length was estimated in conformity to the guidelines. Transit time was calculated using the intersecting tangent method, but in two studies (Vicorder, pOpmètre) best agreement was found using the maximum of the second derivative. Six studies reached the accuracy level 'excellent' defined in the Artery guidelines.
Conclusion:
Method to assess transit time and path length need validation in larger populations. Further studies are required in different risk population to implement clinical applicability of every device.
Related Concept Videos
Velocity and Acceleration of a Wave
The velocity of the particles can be obtained by taking the partial derivative of the position equation with respect to time....
Assessment of apical pulse
Assessing the apical pulse is a critical nursing procedure, particularly indicated for:
Pulse Assessment Sites
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps:
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation
Review and Preview
Percentiles are a type of fractile that partition data into...