Related Experiment Video
Updated: Jan 27, 2026

Treatment of Liver Metastases Using an Internal Target Volume Method for Stereotactic Body Radiotherapy
Published on: May 8, 2018
Distributive quality assurance and delivery of stereotactic ablative radiotherapy treatments amongst beam matched
James Rijken1,2, Hilmar Schachenmayr3, Scott Crowe2,4
1GenesisCare, Flinders Private Hospital, Bedford Park, SA, Australia.
Purpose:
Beam matching occurs on all linacs to some degree and when two are more are matched to each other, patients are able to be transferred between machines. Quality assurance of plans can also be performed "distributively" on any of the matched linacs. The degree to which machines are matched and how this translates to like delivery of plans has been the focus of a number of studies. This concept has not yet been explored for stereotactic techniques which require a higher degree of accuracy. This study proposes beam matching criteria which allows for the distributive delivery and quality assurance of stereotactic body radiotherapy (SBRT) plans.
Method:
Two clinically relevant and complex volumetric modulated arc therapy (VMAT) SBRT spine and lung plans were chosen as benchmarking cases. These were delivered on nine previously beam matched linacs with quality assurance performed through ArcCheck and film exposure in the sagittal plane. Measured doses were compared to their treatment planning system predictions through gamma analysis at a range of criteria.
Results:
Despite differences in beam match parameters and variations in small fields, all nine linacs produced accurate deliveries with a tight deviation in the population sample. Pass rates were well above suggested tolerances at the recommended gamma criterion. Film was able to detect dose errors to a greater degree than ArcCheck.
Conclusion:
Distributive quality assurance and delivery of stereotactic ablative radiotherapy treatments amongst beam matched linacs is certainly feasible provided the linacs are matched to a strict protocol like that suggested in this study and regular quality assurance is performed on the matched fleet. Distributive quality assurance and delivery of SBRT provides the possibility of efficiency gains for physicists as well as treatment staff.
Related Concept Videos
Quality Assurance
Distribution of Stresses in a Narrow Rectangular Beam
Sign Test for Matched Pairs
To conduct the sign test, we first calculate the differences in...
Beams
Based on geometry, beams can be straight, tapered, or curved. Straight beams are the most common type and have a constant cross-section throughout their length. Tapered beams, on the other hand, have a varying cross-section along...
Accelerators
The effectiveness of calcium chloride can...
Linear Circuits