1From Beth Israel Deaconess Medical Center, Boston (J.J.P.); University of Michigan Hospitals, Ann Arbor (G.M.D., S.J.C.), and Spectrum Health Hospitals, Grand Rapids (J.C.H., W.M.) - both in Michigan; Riverside Methodist-Ohio Health, Columbus (S.J.Y.); University of Pittsburgh Medical Center Pinnacle Health, Harrisburg (M.M., H.G.), and the University of Pittsburgh, Pittsburgh (T.G.G.) - both in Pennsylvania; Aurora-Saint Luke's Medical Center, Milwaukee (D.O., T.B.); Houston Methodist Debakey Heart and Vascular Center, Houston (N.S.K., M.J.R.); Abbott Northwestern Hospital (J.A., P.S.) and Medtronic (M.J.B., H.Q.), Minneapolis, Mayo Clinic, Rochester (J.K.O.), and Paradigm Biostatistics, Anoka (A.S.M.) - all in Minnesota; Morton Plant Hospital, Clearwater, FL (J. Rovin); Mount Sinai Health System, New York (D.H.A.), and Saint Francis Hospital, Roslyn (N.R., G.P.) - both in New York; Scripps Clinic and Research Foundation, La Jolla, CA (P.S.T.); University of Kansas Hospital, Kansas City (G.L.Z.); Yale New Haven Hospital, New Haven, CT (J.K.F.); Clinique Pasteur, Toulouse, France (D.T.); Johns Hopkins Hospital, Baltimore (J. Resar); Alfred Hospital, Melbourne, VIC, Australia (A.W.); McGill University Health Centre, Montreal (N.P.); and Winchester Medical Center, Winchester, VA (B.R.).
Background:
Transcatheter aortic-valve replacement (TAVR) is an alternative to surgery in patients with severe aortic stenosis who are at increased risk for death from surgery; less is known about TAVR in low-risk patients.
Methods:
We performed a randomized noninferiority trial in which TAVR with a self-expanding supraannular bioprosthesis was compared with surgical aortic-valve replacement in patients who had severe aortic stenosis and were at low surgical risk. When 850 patients had reached 12-month follow-up, we analyzed data regarding the primary end point, a composite of death or disabling stroke at 24 months, using Bayesian methods.
Results:
Of the 1468 patients who underwent randomization, an attempted TAVR or surgical procedure was performed in 1403. The patients' mean age was 74 years. The 24-month estimated incidence of the primary end point was 5.3% in the TAVR group and 6.7% in the surgery group (difference, -1.4 percentage points; 95% Bayesian credible interval for difference, -4.9 to 2.1; posterior probability of noninferiority >0.999). At 30 days, patients who had undergone TAVR, as compared with surgery, had a lower incidence of disabling stroke (0.5% vs. 1.7%), bleeding complications (2.4% vs. 7.5%), acute kidney injury (0.9% vs. 2.8%), and atrial fibrillation (7.7% vs. 35.4%) and a higher incidence of moderate or severe aortic regurgitation (3.5% vs. 0.5%) and pacemaker implantation (17.4% vs. 6.1%). At 12 months, patients in the TAVR group had lower aortic-valve gradients than those in the surgery group (8.6 mm Hg vs. 11.2 mm Hg) and larger effective orifice areas (2.3 cm2 vs. 2.0 cm2).
Conclusions:
In patients with severe aortic stenosis who were at low surgical risk, TAVR with a self-expanding supraannular bioprosthesis was noninferior to surgery with respect to the composite end point of death or disabling stroke at 24 months. (Funded by Medtronic; ClinicalTrials.gov number, NCT02701283.).
Related Concept Videos
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...