Matrix-assisted ionization mass spectrometry in targeted protein analysis - An initial evaluation
Øystein Skjaervø1, Sarah Trimpin2, Trine Grønhaug Halvorsen1
1Section of Pharmaceutical Chemistry, Department of Pharmacy, University of Oslo, P.O. Box 1068 Blindern, 0316, Oslo, Norway.
Matrix-assisted ionization (MAI) coupled with mass spectrometry (MS) offers a simple method for targeted protein analysis. This technique shows potential for biological samples, reducing matrix effects compared to electrospray ionization (ESI).
Area of Science:
- Analytical Chemistry
- Biochemistry
- Mass Spectrometry
Background:
- Matrix-assisted ionization (MAI) is an emerging ionization technique for mass spectrometry (MS).
- MAI exhibits reduced matrix effects compared to electrospray ionization (ESI).
- These characteristics make MAI suitable for targeted protein analysis in biological samples.
Purpose of the Study:
- To evaluate matrix-assisted ionization coupled with selected reaction monitoring mass spectrometry (MAI-SRM-MS) for targeted protein quantification.
- To assess the applicability of MAI-SRM-MS using a bottom-up proteomic approach with stable isotopic labeled peptides as internal standards.
- To investigate the performance of MAI-SRM-MS in complex biological matrices.
Main Methods:
- Utilized a triple quadrupole mass analyzer with a nanoESI source in selected reaction monitoring (SRM) mode.
- Employed the bottom-up proteomic strategy with stable isotopic labeled peptides as internal standards (IS).
- MAI matrix consisted of 3-nitrobenzonitrile in acetonitrile, mixed with aqueous samples at a 1:3 ratio.
Main Results:
- Established SRM settings for electrospray ionization (ESI-SRM-MS) were transferable to MAI-SRM-MS for most peptides.
- Internal standards effectively corrected quantification across various digest levels (RSD ≤28%) and in complex matrices like diluted urine and serum.
- Sufficient dilution was necessary to mitigate signal suppression from biological matrices.
Conclusions:
- MAI-SRM-MS demonstrates significant potential for simple, rapid, and user-friendly targeted protein determination via the bottom-up approach.
- Further validation is required before clinical implementation of MAI-SRM-MS.
- The technique's robustness in handling biological samples warrants continued investigation.
More Related Videos
10:47Whole-body Mass Spectrometry Imaging by Infrared Matrix-assisted Laser Desorption Electrospray Ionization IR-MALDESI
Published on: March 24, 2016
12:19Uracil-DNA Glycosylase Assay by Matrix-assisted Laser Desorption/Ionization Time-of-flight Mass Spectrometry Analysis
Published on: April 22, 2022
Related Concept Videos
Chemical Ionization (CI) Mass Spectrometry
Matrix-Assisted Laser Desorption Ionization (MALDI)
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
Tandem Mass Spectrometry
Mass Spectrometry: Overview