An experimental method to directly measure DQE[Formula: see text] at k = 0 for 2D x-ray imaging systems
Xu Ji1, Mang Feng1, Ran Zhang1
1Department of Medical Physics, University of Wisconsin-Madison, 1111 Highland Avenue, Madison, WI 53705, United States of America.
Abstract:
The zero-frequency detective quantum efficiency (DQE), viz., DQE0, is defined as the ratio between output and input squared signal-to-noise ratio of an imaging system. In 1963, R. Shaw applied Fourier analysis to generalize DQE0 to the frequency-dependent DQE, i.e. DQE[Formula: see text]. Under conditions specified by Shaw, DQE[Formula: see text] is the same as DQE0 at k = 0. The experimental measurement of DQE[Formula: see text] involves the measurement of system modulation transfer function (MTF) and noise power spectrum (NPS). Although the measurement of MTF is straightforward, the experimental measurements of NPS[Formula: see text] encountered several challenges. As a result, some experimental methods may yield a nonphysical NPS value at k = 0, which makes the measured DQE(k)| k=0 deviate from the true zero-frequency DQE. This work presents new results from three aspects: 1) system drift is a significant error source when a large number of independent image acquisitions are involved in measuring NPS and DQE; 2) a cascaded systems analysis shows that the drift induces a global positive offset to the measured autocovariance function, and the offset is quantitatively related to the NPS error at k = 0; 3) based on the measured autocovariance data, drift-induced offset can be estimated, so that errors in the measured NPS(k)| k=0 and DQE(k)| k=0 can be corrected. Both numerical simulations with known ground truth for DQE0 and experimental studies were performed to validate the proposed measurement method. The results demonstrated that the method mitigates the undesirable influence of system drift in DQE(k)| k=0 and DQE0, allowing the measured values consistent with the classical definition of zero-frequency DQE.
More Related Videos
Related Concept Videos
Experimental Determination of Chemical Formula
Chemical Formulas
X-ray Imaging
Ionic Compounds: Formulas and Nomenclature
Molecular Compounds: Formulas and Nomenclature
Formula Mass and Mole Concepts of Compounds